Abstract
In the study, novel blood pressure estimation method was proposed to improve the accuracy of oscillometric method. The proposed algorithm estimated the blood pressure by comparing and analyzing the point variation aspect of dicrotic notch on pulsating waveform during each cardiac cycle. The waveforms of each cardiac cycle were extracted by maximum points. The extracted pulsating waveforms were applied by re-sampling, end-matching, and normalization. The systolic and diastolic blood pressures were estimated by point variation aspect of dicrotic notch. The blood pressures, which were estimated from proposed algorithm, were compared and analyzed by blood pressures from oscillometric methods and auscultation. The systolic blood pressure from oscillometric methods were +0.88 mmHg more than proposed algorithm, and 1.875 less than the diastolic blood pressures from proposed algorithm. The systolic and diastolic blood pressures from auscultation were 2.89 mmHg and 3.44 mmHg less than the blood pressures from proposed algorithm. As the errors between blood pressures from proposed algorithm, oscillometric method and auscultation were less than 5 mmHg, the proposed algorithm was effective.