Abstract
The patients who had been operated total mastectomy are different from general women in their breasts thickness due to breast surgery. As a result, digital chest image from total mastectomy patients will be different attenuation. The main objective for this study is to show that a proper Ion chamber standard combination measuring MTF which is objective basis for Digital image, when be x-ray for total mastectomy patient. We have designed the unique number that shown Left is 1, Right is 2, Center is 3 and have put the edge phantom on detector ion chamber. Lastly, we have obtained experiment images. The evaluations of all image quality have measured by 50% MTF of spatial resolution and absorption dose using Matlab(R2007a). The result showed that average exposure condition, MTF value, absorption dose for 1+3 and 2+3 combinations were 2.745 mAs, 1.925 lp/mm, 0.688 mGy. Consequently, that showed high MTF, DQE and low dose than other combinations. Therefore, a proper changes of ion chambers are able to improve image quality and to reduce radiation exposure when be X-ray for total mastectomy patients. Also, it will be possible to standard for application chamber combination and utilization on clinical detection.
변형근치유방절제술을 시행한 환자는 일반 여성에 비해 수술로 인한 양측 유방의 두께 차이가 나타나게 되고 이에 따른 감약 차가 생기게 된다. 이에 본 연구는 디지털 영상 특성평가의 객관적인 기준이 되는 MTF를 측정하여 변형근치유방절제술을 시행한 환자들의 흉부촬영에서 적합한 Ion chamber 조합의 기준을 제시하고자 하였다. 디텍터에 위치한 chamber에 좌측 1, 우측 2, 가운데 3으로 번호를 지정하고 엣지팬텀을 부착하여 실험 영상을 획득하였다. 획득한 영상의 해상력 특성 평가는 Matlab(R2007a)를 이용하여 50% MTF 공간주파수의 해상력 평가와 흡수선량을 측정하였다. 실험결과에서 1+3과 2+3조합의 평균 노출조건, 공간주파수, 흡수선량은 각각 2.745 mAs, 1.925 lp/mm, 0.668 mGy로 다른 조합과 비교하여 적은 선량으로 높은 해상력 특성과 높은 DQE를 나타내었다. 따라서 변형근치유방절제술을 시행한 환자의 디지털 흉부영상의 촬영에서는 본 연구에서 제시하는 chamber 조합은 피폭선량 감소와 영상 화질의 향상에 대한 보상이 가능하여 향후 임상에서 적용 및 활용 가능한 chamber 기준이 될 것으로 사료된다.