This paper proposes a thermal peripheral blood flowmeter integrated with a force sensor that is capable of contact force compensation. We fabricate this blood flowmeter using a nickel RTD (resistance temperature detector) and piezoresistive force sensor by using microfabrication technology. In an experiment, we obtained a decreasing trend for the blood flow under an increasing contact force with a linear tendency of 31.7%/N. We then performed a compensation process based on this obtained trend. As a result, the maximum variance in the blood flow at 1-3N was 9.8%. Thus we achieved consistent blood flow measurement independent of the contact force. In this work, we verified that the thermal blood flowmeter integrated with a force sensor has the ability to accurately measure the blood flow independent of the contact force.
본 논문은 접촉력 보정이 가능한 혈류량 측정기를 제안한다. 기존 혈류량 측정기는 접촉력의 영향을 최소화하기 위해 혈류량 센서를 피부에 고정했으나 이런 점은 전자기기에 적용이 용이하지 않다는 단점이 있다. 이에 본 연구에서는 혈류량 센서에 힘 센서를 집적하여 혈류량과 접촉력의 동시 측정을 통해 접촉력에 따른 혈류량 변화 오차를 보정하여 접촉력에 무관한 혈류량 측정이 가능한 소자 및 방법을 제안한다. 제작된 혈류량 측정기의 성능 분석 결과, 접촉력에 따른 혈류량 변화는 31.7%/N 의 선형적 감소경향임을 확인하였고, 이를 이용하여 접촉력에 따른 혈류량 오차를 보정한 결과, 접촉력 1~3N에서 최대 편차가 9.8%로 나타나 다른 접촉력 조건에서도 일정한 혈류량 측정이 가능함을 보였다. 제안한 소자는 접촉력에 무관하게 정확한 혈류량을 측정할 수 있어, 접촉 가능한 전자기기에의 적용이 용이하다.