Influence of the implant abutment types and the dynamic loading on initial screw loosening

  • Kim, Eun-Sook (Department of Prosthodontics, College of Dentistry, Dankook University) ;
  • Shin, Soo-Yeon (Department of Prosthodontics, College of Dentistry, Dankook University)
  • Received : 2012.06.25
  • Accepted : 2012.11.19
  • Published : 2013.02.28

Abstract

PURPOSE. This study examined the effects of the abutment types and dynamic loading on the stability of implant prostheses with three types of implant abutments prepared using different fabrication methods by measuring removal torque both before and after dynamic loading. MATERIALS AND METHODS. Three groups of abutments were produced using different types of fabrication methods; stock abutment, gold cast abutment, and CAD/CAM custom abutment. A customized jig was fabricated to apply the load at $30^{\circ}$ to the long axis. The implant fixtures were fixed to the jig, and connected to the abutments with a 30 Ncm tightening torque. A sine curved dynamic load was applied for $10^5$ cycles between 25 and 250 N at 14 Hz. Removal torque before loading and after loading were evaluated. The SPSS was used for statistical analysis of the results. A Kruskal-Wallis test was performed to compare screw loosening between the abutment systems. A Wilcoxon signed-rank test was performed to compare screw loosening between before and after loading in each group (${\alpha}$=0.05). RESULTS. Removal torque value before loading and after loading was the highest in stock abutment, which was then followed by gold cast abutment and CAD/CAM custom abutment, but there were no significant differences. CONCLUSION. The abutment types did not have a significant influence on short term screw loosening. On the other hand, after $10^5$ cycles dynamic loading, CAD/CAM custom abutment affected the initial screw loosening, but stock abutment and gold cast abutment did not.

Keywords

References

  1. Schmitt A, Zarb GA. The longitudinal clinical effectiveness of osseointegrated dental implants for single-tooth replacement. Int J Prosthodont 1993;6:197-202.
  2. Wannfors K, Smedberg JI. A prospective clinical evaluation of different single-tooth restoration designs on osseointegrated implants. A 3-year follow-up of Brånemark implants. Clin Oral Implants Res 1999;10:453-8. https://doi.org/10.1034/j.1600-0501.1999.100603.x
  3. Laney WR, Jemt T, Harris D, Henry PJ, Krogh PH, Polizzi G, Zarb GA, Herrmann I. Osseointegrated implants for single-tooth replacement: progress report from a multicenter prospective study after 3 years. Int J Oral Maxillofac Implants 1994;9:49-54.
  4. Jemt T, Pettersson P. A 3-year follow-up study on single implant treatment. J Dent 1993;21:203-8. https://doi.org/10.1016/0300-5712(93)90127-C
  5. Kang YM, Lim JH, Cho IH. A study on the abutment screw loosening of dental implants. J Korean Acad Prosthodont 1996;34:1-14.
  6. Lazzara R, Siddiqui AA, Binon P, Feldman SA, Weiner R, Phillips R, Gonshor A. Retrospective multicenter analysis of 3i endosseous dental implants placed over a five-year period. Clin Oral Implants Res 1996;7:73-83. https://doi.org/10.1034/j.1600-0501.1996.070109.x
  7. Jörnéus L, Jemt T, Carlsson L. Loads and designs of screw joints for single crowns supported by osseointegrated implants. Int J Oral Maxillofac Implants 1992;7:353-9.
  8. Burguete RL, Johns RB, King T, Patterson EA. Tightening characteristics for screwed joints in osseointegrated dental implants. J Prosthet Dent 1994;71:592-9. https://doi.org/10.1016/0022-3913(94)90443-X
  9. Haack JE, Sakaguchi RL, Sun T, Coffey JP. Elongation and preload stress in dental implant abutment screws. Int J Oral Maxillofac Implants 1995;10:529-36.
  10. Martin WC, Woody RD, Miller BH, Miller AW. Implant abutment screw rotations and preloads for four different screw materials and surfaces. J Prosthet Dent 2001;86:24-32. https://doi.org/10.1067/mpr.2001.116230
  11. Cibirka RM, Nelson SK, Lang BR, Rueggeberg FA. Examination of the implant-abutment interface after fatigue testing. J Prosthet Dent 2001;85:268-75. https://doi.org/10.1067/mpr.2001.114266
  12. Sawase T, Wennerberg A, Hallgren C, Albrektsson T, Baba K. Chemical and topographical surface analysis of five different implant abutments. Clin Oral Implants Res 2000;11: 44-50. https://doi.org/10.1034/j.1600-0501.2000.011001044.x
  13. Heydecke G, Sierraalta M, Razzoog ME. Evolution and use of aluminum oxide single-tooth implant abutments: a short review and presentation of two cases. Int J Prosthodont 2002; 15:488-93.
  14. Wood MR, Vermilyea SG; Committee on Research in Fixed Prosthodontics of the Academy of Fixed Prosthodontics. A review of selected dental literature on evidence-based treatment planning for dental implants: report of the Committee on Research in Fixed Prosthodontics of the Academy of Fixed Prosthodontics. J Prosthet Dent 2004; 92:447-62. https://doi.org/10.1016/j.prosdent.2004.08.003
  15. Priest G. Virtual-designed and computer-milled implant abutments. J Oral Maxillofac Surg 2005;63:22-32.
  16. Ortorp A, Jemt T, Bäck T, Jälevik T. Comparisons of precision of fit between cast and CNC-milled titanium implant frameworks for the edentulous mandible. Int J Prosthodont 2003;16:194-200.
  17. Al-Fadda SA, Zarb GA, Finer Y. A comparison of the accuracy of fit of 2 methods for fabricating implant-prosthodontic frameworks. Int J Prosthodont 2007;20:125-31.
  18. Oh BD, Choi YS, Shin SY. The effect of types of abutment and dynamic loading on microgap between implant fixture and abutment. J Dent Rehabil Appl Sci 2010;26:389-403.
  19. Kapos T, Ashy LM, Gallucci GO, Weber HP, Wismeijer D. Computer-aided design and computer-assisted manufacturing in prosthetic implant dentistry. Int J Oral Maxillofac Implants 2009;24:110-7.
  20. Schwarz MS. Mechanical complications of dental implants. Clin Oral Implants Res 2000;11:156-8. https://doi.org/10.1034/j.1600-0501.2000.011S1156.x
  21. Binon PP. Evaluation of machining accuracy and consistency of selected implants, standard abutments, and laboratory analogs. Int J Prosthodont 1995;8:162-78.
  22. Merz BR, Hunenbart S, Belser UC. Mechanics of the implant- abutment connection: an 8-degree taper compared to a butt joint connection. Int J Oral Maxillofac Implants 2000;15:519-26.
  23. Norton MR. An in vitro evaluation of the strength of an internal conical interface compared to a butt joint interface in implant design. Clin Oral Implants Res 1997;8:290-8. https://doi.org/10.1034/j.1600-0501.1997.080407.x
  24. Weiss EI, Kozak D, Gross MD. Effect of repeated closures on opening torque values in seven abutment-implant systems. J Prosthet Dent 2000;84:194-9. https://doi.org/10.1067/mpr.2000.108069
  25. Norton MR. Assessment of cold welding properties of the internal conical interface of two commercially available implant systems. J Prosthet Dent 1999;81:159-66. https://doi.org/10.1016/S0022-3913(99)70243-X
  26. Balfour A, O'Brien GR. Comparative study of antirotational single tooth abutments. J Prosthet Dent 1995;73:36-43. https://doi.org/10.1016/S0022-3913(05)80270-7
  27. Theoharidou A, Petridis HP, Tzannas K, Garefis P. Abutment screw loosening in single-implant restorations: a systematic review. Int J Oral Maxillofac Implants 2008;23: 681-90.
  28. Jaarda MJ, Razzoog ME, Gratton DG. Effect of preload torque on the ultimate tensile strength of implant prosthetic retaining screws. Implant Dent 1994;3:17-21. https://doi.org/10.1097/00008505-199404000-00002
  29. Siamos G, Winkler S, Boberick KG. Relationship between implant preload and screw loosening on implant-supported prostheses. J Oral Implantol 2002;28:67-73. https://doi.org/10.1563/1548-1336(2002)028<0067:TRBIPA>2.3.CO;2
  30. Tan KB, Nicholls JI. Implant-abutment screw joint preload of 7 hex-top abutment systems. Int J Oral Maxillofac Implants 2001;16:367-77.
  31. Binon PP, McHugh MJ. The effect of eliminating implant/ abutment rotational misfit on screw joint stability. Int J Prosthodont 1996;9:511-9.
  32. Lee J, Kim YS, Kim CW, Han JS. Wave analysis of implant screw loosening using an air cylindrical cyclic loading device. J Prosthet Dent 2002;88:402-8. https://doi.org/10.1067/mpr.2002.128099
  33. Vigolo P, Fonzi F, Majzoub Z, Cordioli G. Evaluation of gold-machined UCLA-type abutments and CAD/CAM titanium abutments with hexagonal external connection and with internal connection. Int J Oral Maxillofac Implants 2008;23:247-52.
  34. Abduo J, Lyons K, Bennani V, Waddell N, Swain M. Fit of screw-retained fixed implant frameworks fabricated by different methods: a systematic review. Int J Prosthodont 2011;24: 207-20.
  35. Bates JF, Stafford GD, Harrison A. Masticatory function - a review of the literature. III. Masticatory performance and efficiency. J Oral Rehabil 1976;3:57-67. https://doi.org/10.1111/j.1365-2842.1976.tb00929.x
  36. Mohl ND, Zarb GA, Carlsson GE, Rugh JD. A textbook of occlusion. Quintessence pub; 1998. p. 148-52.
  37. Gibbs CH, Mahan PE, Mauderli A, Lundeen HC, Walsh EK. Limits of human bite strength. J Prosthet Dent 1986; 56:226-9. https://doi.org/10.1016/0022-3913(86)90480-4
  38. Bae BR, Choi YS, Cho IH. Influence of implant abutment systems on detorque value and screw joint stability. J Dent Rehabil Appl Sci 2010;26:97-109.

Cited by

  1. Comparative Analysis of Screw Loosening With Prefabricated Abutments and Customized CAD/CAM Abutments vol.25, pp.6, 2016, https://doi.org/10.1097/ID.0000000000000481
  2. Effects of grooved abutment on stability of implant abutment screw vol.54, pp.4, 2016, https://doi.org/10.4047/jkap.2016.54.4.387
  3. Effect of abutment screw length and cyclic loading on removal torque in external and internal hex implants vol.8, pp.1, 2016, https://doi.org/10.4047/jap.2016.8.1.62
  4. In Vitro Laser Treatment Platform Construction with Dental Implant Thread Surface on Bacterial Adhesion for Peri-Implantitis vol.2017, pp.2314-6141, 2017, https://doi.org/10.1155/2017/4732302
  5. Influence of Abutment Surface Treatments on Screw Loosening of Morse Taper Implants vol.26, pp.5, 2017, https://doi.org/10.1097/ID.0000000000000622
  6. Loosening of the fixing screw in single implant crowns: predisposing factors, prevention and treatment options vol.29, pp.4, 2017, https://doi.org/10.1111/jerd.12303
  7. Marginal Vertical Fit along the Implant-Abutment Interface: A Microscope Qualitative Analysis vol.4, pp.3, 2016, https://doi.org/10.3390/dj4030031
  8. Mechanical Outcomes, Microleakage, and Marginal Accuracy at the Implant-Abutment Interface of Original versus Nonoriginal Implant Abutments: A Systematic Review of In Vitro Studies vol.2018, pp.2314-6141, 2018, https://doi.org/10.1155/2018/2958982
  9. Load Fatigue Performance Evaluation on Two Internal Tapered Abutment-Implant Connection Implants Under Different Screw Tightening Torques vol.43, pp.2, 2013, https://doi.org/10.1563/aaid-joi-d-16-00129
  10. Effect of Abutment Connection Type and Cyclic Loading on Removal Torque Value vol.4, pp.1, 2013, https://doi.org/10.29252/jrdms.4.1.36
  11. Mechanical Complications of Posterior Implant Restorations using CAD/CAM Titanium Customized Abutments: A Retrospective Study vol.23, pp.3, 2013, https://doi.org/10.32542/implantology.2019012
  12. Custom abutments on tilted implants in the maxilla: A clinical report vol.17, pp.4, 2013, https://doi.org/10.4103/1735-3327.292065
  13. Hybrid-abutment-restoration: effect of material type on torque maintenance and fracture resistance after thermal aging vol.6, pp.None, 2013, https://doi.org/10.1186/s40729-020-00220-y
  14. Effect of Serrating Abutment-Implant Mating Surface on Torque Stability of Implant-Abutment Connection, Before and After Cyclic Loading vol.46, pp.6, 2013, https://doi.org/10.1563/aaid-joi-d-19-00029
  15. Investigation of the Effect of Abutment Angle Tolerance on the Stress Created in the Fixture and Screw in Dental Implants Using Finite Element Analysis vol.51, pp.None, 2013, https://doi.org/10.4028/www.scientific.net/jbbbe.51.63
  16. Oral reconstruction with hybrid implant‐supported fixed prostheses in cases of mandibular defect using two methods along with 3 years of follow‐up: A case report vol.9, pp.9, 2013, https://doi.org/10.1002/ccr3.4854
  17. Clinical outcomes of single implant supported crowns versus 3-unit implant-supported fixed dental prostheses in Dubai Health Authority: a retrospective study vol.21, pp.1, 2021, https://doi.org/10.1186/s12903-021-01530-2