Calibration of Optically Stimulated Luminescent nanoDot Dosimeter for 6 MV Photon Beam

6 MV 광자 빔에 대한 광자극형광나노닷선량계의 교정

  • Kim, Jongeon (Department of Radiological Science, Kaya University) ;
  • Kim, Seonghu (Department of Radiology, Gyeongsang National University Hospital) ;
  • Lee, Hyoyeong (Department of Nuclear Medicine, Pusan National University Yangsan Hospital)
  • 김종언 (가야대학교 방사선학과) ;
  • 김성후 (경상대학교병원 영상의학과) ;
  • 이효영 (부산대학교병원 핵의학과)
  • Received : 2013.01.20
  • Accepted : 2013.02.22
  • Published : 2013.02.28

Abstract

The purpose of this study is to investigate the calibration of an optically stimulated luminescent nanoDot dosimeter(OSLnD) to 6 MV photon beam. Dose ranges of the calibration of linear and non-linear from the analysis of dose response of the OSLnD were decided. To evaluate the accuracy of calibration equation and the calibration, the sets of the calibration and quality control dosimeter were used to make. The calibrations were performed by the linear and the non-linear in the dose range of 0~300 cGy and 20~1300 cGy, respectively. The errors of the calibration were acquired less than 0.1% respectively from the measurement of the quality control dosimeters for the calibration of linear and the non-linear. This study provides the calibration equation of the OSLnD to the 6 MV photon beam.

이 연구의 목적은 6 MV 광자 빔에 대한 광자극형광나노닷선량계(OSLnD)의 교정을 조사하는 데 있다. OSLnD의 선량반응 분석으로부터 선형 및 비선형 교정의 선량범위들은 결정을 하였다. 교정 및 교정식의 정확성을 평가하기 위하여 교정 및 품질관리선량계의 세트들은 만들어서 사용하였다. 교정들은 각각 0~300 cGy 및 20~1300 cGy 선량범위에서 선형 및 비선형적으로 수행하였다. 교정의 오차들은 선형 및 비선형 교정에 대하여 품질관리선량계들의 측정으로부터 각각 0.1% 이하로 얻었다. 이 연구는 6 MV 광자 빔에 대한 OSLnD의 교정식을 제공한다.

Keywords

References

  1. D.W. Kim, W.K. Chung, D.O. Shin, M.G. Yoon and U.J. Hwang,"Dose response of commercially available optically stimulated luminescent detector, Al2O3:C for megavoltage photons and electrons", Radiation Protection Dosimetry, pp.1-8, 2011.
  2. P.A. Jursinic, "Changes in optically stimulated luminescent dosimeter(OSLD) dosimetric characteristics with accumulated dose", Med. Phys, Vol. 27, No. 1, pp.132-140, 2010.
  3. P.A. Jursinic, "Characterization of optically stimulated luminescent dosimeters, OSLDs, for clinical dosimetric measurements", Med. Phys, Vol. 34, No. 12, pp.4594-4604, 2007. https://doi.org/10.1118/1.2804555
  4. M.S. Akselrod, A.C. Lucas, J.C. Polf and S.W. Mckeever, "Optically stimulated luminescence of Al2O3:C", Radiat. Meas, Vol. 29, pp.391-399, 1998. https://doi.org/10.1016/S1350-4487(98)00061-4
  5. microStar user manual, Landauer Inc., 2009.
  6. Frequently asked questions about landaue's optically stimulated luminescence(OSL) detectors, Landauer Inc
  7. A. Viamonte, L.A.R. Darosa, L.A. Buckley, A. Cherpak and J.E. Cygler, "Radiotherapy dosimetry using a commercial OSL system", Med. Phys, Vol. 35, No. 4, pp.1261-1266, 2008. https://doi.org/10.1118/1.2841940
  8. V. Schembri and B.J.M. Heijmen, "Optically stimulated luminescence(OSL) of carbon-doped aluminum oxide ( Al2O3:C) for film dosimetry in radiotherapy", Med. Phys, Vol. 34, No. 6, pp.2113-2118, 2007. https://doi.org/10.1118/1.2737160
  9. M.S. Yukihara, V.H. Whitley, S.W.S. Mckeever, A.E. Akselrod and M.S. Akselrod, "Effects of high-dose irradiation on the optically stimulaetd luminescence of Al2O3:C", Radiat. Meas, Vol. 38, pp.317-330, 2004. https://doi.org/10.1016/j.radmeas.2004.01.033
  10. E.G.Yukihara, G. Mardirossian, M. Mirzasadeghi, S. Guduru and S. Ahmad, Evaluation of Al2O3:C optically stimulated luminescence(OSL) dosimeters for passive dosimetry of high-energy photon and electron beams in radiotherapy, Med. Phys, Vol. 35, No. 1, pp.260-269, 2008. https://doi.org/10.1118/1.2816106

Cited by

  1. Measurement of Dose outside a 6 MV Field Edge Using Optically Stimulated Luminescent Nano Dot Dosimeters vol.8, pp.7, 2014, https://doi.org/10.7742/jksr.2014.8.7.449