DOI QR코드

DOI QR Code

Design of a Four-axis Force/Moment Sensor for Measuring the Applied Force to Wrist

손목에 가해지는 힘측정을 위한 4축 힘/모멘트센서 설계

  • Hong, Tae-Kyung (Department of Control & Instrumentation Engineering, Gyeongsang National University) ;
  • Kim, Gab-Soon (Department of Control & Instrumentation Engineering, Gyeongsang National University)
  • 홍태경 (경상대학교 제어계측공학과, ERI) ;
  • 김갑순 (경상대학교 제어계측공학과, ERI)
  • Received : 2013.05.28
  • Accepted : 2013.08.22
  • Published : 2013.11.01

Abstract

Patients have the paralysis of their wrists, and can't use of their wrists freely. But their wrists can be recovered by wrist-bending rehabilitation exercise. Professional rehabilitation therapeutists exercise the wrists of patients in hospital. But the wrists of patients have not exercised enough for the rehabilitation, because the therapeutists are much less than patients in number. Therefore, the wrist rehabilitation robot should be developed, and it have to measure the applied force to the patients' wrists for their safety. In this paper, the four-axis force/moment sensor was designed for the wrist rehabilitation robot. As a test results, the interference error of the four-axis force/moment sensor was less than 0.91%. It is thought that the sensor can be used to measure the applied force to the patients' wrists.

Keywords

References

  1. P. R. Culmer, A. E. Jackson, S. Makower, R. Richardson, J. A. Cozens, M. C. Levesley, and B. B. Bhakta, "A control strategy for upper limb robotic rehabilitation with a dual robot system," IEEE/ASME Transactions on Mechatronics, vol. 15, no. 4, pp. 575-585, Aug. 2010. https://doi.org/10.1109/TMECH.2009.2030796
  2. L. Pan, A. Song, G. Xu, H. Li, H. Zeng, and B. Xu, "Safety supervisory strategy for an upper-limb rehabilitation robot based on impedance control," International Journal of Advanced Robotic Systems, vol. 10, pp. 1-12, 2013.
  3. H. C. Kim, L. M. Miller, I. Fedulow, M. Simkins, G. M. Abrams, N. Byl, and J. Rosen, "Kinematic data analysis for post-stroke patients following bilateral versus unilateral rehabilitation with an upper limb wearable robotic system," IEEE Transactions on Neural Systems and Rehabilitation Engineering, vol. 21, no. 2, pp. 153-164, 2013. https://doi.org/10.1109/TNSRE.2012.2207462
  4. Y. G. Kim and G. S. Kim, "Design of six-axis force/moment sensor for ankle-rehabilitation robot," Journal of Institute of Control, Robotics and Systems (in Korean), vol. 19, no. 4, pp. 357-363, 2013. https://doi.org/10.5302/J.ICROS.2013.13.1865
  5. G. S. Kim, "Development of cylindrical-type finger force measuring system using two-axis force/moment sensor and its characteristic evaluation," Journal of Institute of Control, Robotics and Systems (in Korean), vol. 17, no. 5, pp. 484-489, 2011. https://doi.org/10.5302/J.ICROS.2011.17.5.484
  6. G. S. Kim and J. J. Park, "Development of the 6-axis force/moment sensor for an intelligent robot's gripper," Sensors and Actuators A, vol. 118, pp. 127-134, 2005. https://doi.org/10.1016/j.sna.2004.07.013
  7. ATI Industrial Automation, "Multi-axis forcre/torque sensor," Ati Industrial Automation, pp. 4-45, 2005.
  8. G. S. Kim and J. W. Yoon, "Development of calibration system for multi-axis force/moment sensor and its uncertainty evaluation," KSPE, vol. 24, no. 10, pp. 91-98, 2007.

Cited by

  1. Rapid Prototyping of Head-of-Bed Angle Measurement System using Open-Source Hardware vol.21, pp.11, 2015, https://doi.org/10.5302/J.ICROS.2015.15.0172
  2. Mechatronic Device for Elbow Rehabilitation vol.53, pp.2261-236X, 2016, https://doi.org/10.1051/matecconf/20165301056
  3. Study, Design and Preliminary Tests of an Automatic Device for Elbow Rehabilitation vol.783, pp.1662-7482, 2015, https://doi.org/10.4028/www.scientific.net/AMM.783.1