Due to the stringent drinking water quality, formaldehyde will be included in Korean drinking water standard from year 2014. However, its standard analytical method has not yet been established. This study compares two analytical methods, DNPH-LC and ACFAS with respect to their analysis principles, Method Detection Limit (MDL), Limit Of Quantitation(LOQ), precision, accuracy, reproducibility, convenience, number of samples analyzed per hour and analysis cost. These methods measure absorption intensity at 360 nm by using HPLC after DNPH-derivatization (DNPH-LC) and at 410 nm by using Automated Continuous Flow Absorption Spectrophotometer (ACFAS), respectively. Reproducibility was tested by repeating the analysis 7 times using a standard solution for each method. For DNPH-LC method, MDL was $0.5{\mu}g/L$, LOQ was $1.58{\mu}g/L$ with standard deviation of $0.16{\mu}g/L$. For ACFAS method, they were $0.27{\mu}g/L$, $0.85{\mu}g/L$L with standard deviation of $0.09{\mu}g/L$, respectively. Both methods satisfied the requirement set by the Korean drinking water quality standard. Complexity of sample pretreatment procedure for DNPH-LC method may cause large error and, consequently, the analytical result will depend on the level of skill of analyst. In contrast, ACFAS method which used only one reagent equipped with an automated injection device showed little analytical error. It costs about $5.00 and $1.00 for one sample to analyze by the DNPH-LC method and the ACFAS method, respectively. Compared to the DNPH-LC method, ACFAS method provided more reliable analytical results. In terms of convenience, easiness and analytical cost, ACFAS method was demonstrated to be superior to the DNPH-LC method. The results of this study suggested that the ACFAS method could be adapted as a proper method for determining formaldehyde content in drinking water.

Keywords

References

  1. Young -Hee Kang (2008) Encyclopedia of Life Science, Academybooks, Seoul.
  2. Sung- Ok Kim, Kyung-Min Jo, Jae-Woon Lee, Hag-Seong Kim (2010) Natural Formation of Formaldehyde in an Anaerobic Water Environment, Fall Association Conference Proceedings of the Korean Society of Water and Wastewer, Nov. 10-11, Gyeonggi, pp.315-316.
  3. Eun-S un Kim, Sun-Hee Ahn, Kang-Yoon Kim, Ho-Choon Choi (2000) Compariosn of Formaldehyde determinatin in air Using Spectrophotometer, Gas Chromatograph, Liquid Chromatograph, Industrial Health, Vol.152 No.0, pp.28-36.
  4. Joung- Hwa Kim, Eun-Suk Choi, Sun-Ku Park, Jun-Hwan Kim , Youn-Tack Lim, Hee- Su Pyo (2004) A Study on the Unregulated Contaminants Management of Drinking Water, Fall Association Conference Proceedings of the Korean Society of Water and Wastewer, Nov.3-4, Daejeon, pp.C-72-73.
  5. National Institute of Environmental Research, (2010) A Study on securing safety of bottled water, pp.16-17.
  6. Dong- Hwan Jeong, Jin-Soo Shin, Ki-Sik Shin, Jae-Hoon Kim, Yong-Seok Kim, Doug- Hee Rhew (2013) A Study on the Determination of Formaldehyde Effluent Limitation in the Industrial Wastewater, Journal of Environmental Impact Assessment, Vol. 22, No. 3 pp.203-217.
  7. Youn- Hee Lee, Jun-Bae Lee, Ju-Hyun Park, Ja-Yoon Choi, Kyung-Hee Ahn, Hye-Sil Ahn, Oh-Sang Kwon (2011) Concentration of Formaldehyde and Acetaldehyde depending on the Time of Storage into Mineral Water, Korean J Environ Agriculture, vol. 30, No. 3, pp. 281-287 https://doi.org/10.5338/KJEA.2011.30.3.281
  8. Seon- Ju Bae, Sang-Ki Chun, In-Chul Yun, Geon-Ho Yi, Dong-Jin Kim (2013) The effects of interfering as acetaldehyde during formaldehyde analysis in drinking Water by ACFAS, 51th Fall Conference Proceedings of The Korean Society of Analytical Sciences, Nov. 21-22, Jeju, p. 214.
  9. Ministry of Environment (2013) Korean Standard Test Methods of Drinking Water (ES 05553.1, ES 05553.2).
  10. American Society for Testing and Materials, ASTM D6303-98. Standard Test Method for Formaldehyde in water.
  11. Japanese Industrial Standards, JIS K0102, Testing methods for industrial wastewater.
  12. Nation al Safty Council, http://www.nsc.org/ news_resources/Resources/Documents/ Formaldehyde.pdf
  13. T. Nash (1953) The colorimetric estimation of formaldehyde by meass of the Hantzsch Reaction, Biochemistry vol. 55, 416-421 https://doi.org/10.1042/bj0550416
  14. US EPA (2009) Drinking water standards and health advisories table, p.3.

Cited by

  1. Effect of Heating Temperature and Time of Coffee Waste on The Adsorptivity of Formaldehyde vol.43, pp.3, 2015, https://doi.org/10.5658/WOOD.2015.43.3.390
  2. 목질 섬유판 제조에 있어 도계부산물인 닭털의 목섬유 부분적 대체화 탐색 vol.56, pp.4, 2013, https://doi.org/10.9713/kcer.2018.56.4.577