DOI QR코드

DOI QR Code

Fatal Subacute Stent Thrombosis Induced by Guidewire Fracture with Retained Filaments in the Coronary Artery

  • Kim, Tae-Jin (Division of Cardiology, Department of Internal Medicine, Haeundae Paik Hospital, Inje University College of Medicine) ;
  • Kim, Jae-Kyun (Division of Cardiology, Department of Internal Medicine, Haeundae Paik Hospital, Inje University College of Medicine) ;
  • Park, Bo-Min (Division of Cardiology, Department of Internal Medicine, Haeundae Paik Hospital, Inje University College of Medicine) ;
  • Song, Pil-Sang (Division of Cardiology, Department of Internal Medicine, Haeundae Paik Hospital, Inje University College of Medicine) ;
  • Kim, Dong-Kie (Division of Cardiology, Department of Internal Medicine, Haeundae Paik Hospital, Inje University College of Medicine) ;
  • Kim, Ki-Hun (Division of Cardiology, Department of Internal Medicine, Haeundae Paik Hospital, Inje University College of Medicine) ;
  • Seol, Sang-Hoon (Division of Cardiology, Department of Internal Medicine, Haeundae Paik Hospital, Inje University College of Medicine) ;
  • Kim, Doo-Il (Division of Cardiology, Department of Internal Medicine, Haeundae Paik Hospital, Inje University College of Medicine)
  • Received : 2013.05.14
  • Accepted : 2013.07.17
  • Published : 2013.11.30

Abstract

During percutaneous coronary intervention, guidewire fractures are very exceptionally encountered in medical practice, but can cause fatal complications such as intracoronary thrombus formation, embolization and perforation. Removal of the remnant segments of guidewire is important for the prognosis. There are several methods being recommended for the treatment of fractured guidewire remnants. However, the best treatment of remnant guidewire filament is still unclear. Herein, we present a case where we did not completely remove remnant guidewire filaments that caused fatal coronary thrombosis.

Keywords

References

  1. Hartzler GO, Rutherford BD, McConahay DR. Retained percutaneous transluminal coronary angioplasty equipment components and their management. Am J Cardiol 1987;60:1260-4. https://doi.org/10.1016/0002-9149(87)90604-7
  2. Khonsari S, Livermore J, Mahrer P, Magnusson P. Fracture and dislodgment of floppy guidewire during percutaneous transluminal coronary angioplasty. Am J Cardiol 1986;58:855-6. https://doi.org/10.1016/0002-9149(86)90370-X
  3. Woodfield SL, Lopez A, Heuser RR. Fracture of coronary guidewire during rotational atherectomy with coronary perforation and tamponade. Cathet Cardiovasc Diagn 1998;44:220-3. https://doi.org/10.1002/(SICI)1097-0304(199806)44:2<220::AID-CCD20>3.0.CO;2-P
  4. Al-Amri HS, AL-Moghairi AM, Calafiore AM. Left main approach for retrieval of retained guidewire fragment. J Card Surg 2012;27:307-8. https://doi.org/10.1111/j.1540-8191.2011.01352.x
  5. van Gaal WJ, Porto I, Banning AP. Guide wire fracture with retained filament in the LAD and aorta. Int J Cardiol 2006;112:e9-11. https://doi.org/10.1016/j.ijcard.2006.01.040
  6. Hong YM, Lee SR. A case of guide wire fracture with remnant filaments in the left anterior descending coronary artery and aorta. Korean Circ J 2010;40:475-7. https://doi.org/10.4070/kcj.2010.40.9.475
  7. Alexiou K, Kappert U, Knaut M, Matschke K, Tugtekin SM. Entrapped coronary catheter remnants and stents: must they be surgically removed? Tex Heart Inst J 2006;33:139-42.
  8. Kim JH, Kim GH, Moon KW. Successful transradial retrieval of broken catheter fragment during transradial coronary angiography. J Invasive Cardiol 2012;24:74-5.
  9. Karabulut A, Daglar E, Cakmak M. Entrapment of hydrophilic coated coronary guidewire tips: which form of management is best? Cardiol J 2010;17:104-8.
  10. Collins N, Horlick E, Dzavik V. Triple wire technique for removal of fractured angioplasty guidewire. J Invasive Cardiol 2007;19:E230-4.
  11. Cho YH, Park S, Kim JS, et al. Rescuing an entrapped guidewire using a Tornus catheter. Circ J 2007;71:1326-7. https://doi.org/10.1253/circj.71.1326
  12. Kim JY, Yoon J, Jung HS, et al. Broken guidewire fragment in the radio- brachial artery during transradial sheath placement: percutaneous retrieval via femoral approach. Yonsei Med J 2005;46:166-8. https://doi.org/10.3349/ymj.2005.46.1.166
  13. Kang JH, Rha SW, Lee DI, et al. Successful retrieval of a fractured and entrapped 0.035-inch terumo wire in the femoral artery using biopsy forceps. Korean Circ J 2012;42:201-4. https://doi.org/10.4070/kcj.2012.42.3.201
  14. Darwazah AK, Abu Sham'a RA, Yassin IH, Islim I. Surgical intervention to remove an entrapped fractured guidewire during angioplasty. J Card Surg 2007;22:526-8. https://doi.org/10.1111/j.1540-8191.2007.00462.x

Cited by

  1. 경피적 관상동맥 확장 성형술 도중 우관상동맥 뒤가쪽가지에서 발생한 유도철사 부러짐 vol.33, pp.1, 2016, https://doi.org/10.12701/yujm.2016.33.1.52
  2. Lost in circulation vol.35, pp.8, 2013, https://doi.org/10.1111/jocs.14821
  3. Device entrapment during percutaneous coronary intervention of chronic total occlusions: incidence and management strategies vol.17, pp.3, 2013, https://doi.org/10.4244/eij-d-20-00781