Abstract
Digital holographic microscopy allows optical path difference measurement. Optical path difference depends on both the refractive index and the morphology of the sample. When interference fringes are very closely spaced, the phase data contain high frequencies where $2{\pi}$ ambiguities cannot be resolved. The immersion testing method, which is a transmission test while the sample is immersed in liquid, is very effective in reducing high frequency fringes in transmission measurements so that large dynamic range testing is possible for a non-null configuration. We developed a digital holographic microscope using liquid that can measure the high numerical aperture aspheric morphology of a sample. This system provides highly precise three-dimensional information on the sample. By improving the experimental method, choosing liquids which have similar refractive index to the sample, we can measure more accurate three-dimensional information on the samples.
디지털 홀로그래피 현미경은 기준광과 시료의 굴절률과 두께에 의한 광경로 차이를 간섭 방법을 이용하여 측정하여 3차원 형상을 측정하는 방법이다. 그리고 광 경로차는 시료의 두께와 시료와 시료 주변의 굴절률 차이에 의존한다. 시료 형상 기울기가 매우 큰 경우에는 광 경로차가 커져 일반적인 간섭 방법으로는 측정이 어렵다. 이런 경우 시료를 액체에 담그어 측정하면 광 경로차가 매우 작아져서 일반적인 간섭 방법을 이용하여 측정할 수 있다. 본 연구에서는 디지털 홀로그램 현미경과 액체를 이용하여 기울기가 큰 비구면 렌즈의 형상을 측정하는 연구를 하였다. 시료의 굴절률과 비슷한 액체를 선택함으로서 비구면 렌즈의 정밀한 3차원 측정이 가능하였다.