Abstract
In this study, based on IEC 60601-2-54 standard load conditions presented in the limited interval over the air kerma at the absolute linearity closely evaluated by measuring the X-ray results were as follows: 10 units targeted all Diagnostic X-ray generating device (DR) presented in the IEC 60601-2-54 standard linearity of air kerma emerged as inappropriate, the general evaluation of the dose linearity from four in the top 50% and 80 kVp % of the two measurement series were as irrelevant all the rest from six of the top tube voltage of 50% and 80% of the two measurement series, appeared in all suitable. Presented in IEC 60601-2-54 standard dose linearity testing and conventional linearity tests showed many differences. IEC 60601-2-54 standard linearity in the proposed international standards of air kerma is the recommendation of the existing dose linearity considerably more feasible, and to quantify the amount of radiation as the standard suggested by the standard IEC 60601-2-54 air kerma of a diagnostic X-ray imaging device linearity performance management is considered key elements in the critical appraisal.
본 연구에서는 IEC 60601-2-54에서 제시한 표준을 근거로 부하조건에서의 제한된 간격으로 걸친 공기커마를 절대측정하여 X선의 직선성을 면밀히 평가하여 다음과 같은 결과를 얻었다. 10대를 대상으로 한 모든 진단용 X선 발생장치(DR)는 IEC 60601-2-54 표준에서 제시한 공기커마의 직선성이 부적합으로 나타났고, 일반적인 조사선량의 직선성 평가방법에서는 4대에서 최고 관전압의 50%와 80%의 두 개 측정 계열 모두 부적합으로 나타났으며, 나머지 6대에서는 최고 관전압의 50%와 80%의 두 개 측정 계열 모두에서 적합으로 나타났다. IEC 60601-2-54 표준에서 제시한 직선성 시험과 기존의 조사선량 직선성 시험이 많은 차이를 보였다. IEC 60601-2-54 표준+에서 제시한 공기커마의 직선성은 국제표준의 권고사항이므로 기존의 조사선량 직선성 보다 상당히 타당성이 있고, 방사선량을 정량화하기 위한 표준으로 IEC 60601-2-54 표준에서 제시한 공기커마의 직선성이 진단용 X선 영상장치 성능관리에 중요한 평가 요소라 사료된다.