Abstract
A polymer gel dosimeter was fabricated. A 3-dimensional dosimetry experiment was performed in the small field of the photon of the cyberknife. The dosimeter was installed in a head and neck phantom. It was manufactured from the acrylic and it was used in dosimetry. By using the head and neck CT protocol of the CyberKnife system, CT images of the head and neck phantom were obtained and delivered to the treatment planning system. The irradiation to the dosimeter in the treatment planning was performed, and then, the image was obtained by using 3.0T magnetic resonance imaging (MRI) after 24 hours. The dose distribution of the phantom was analyzed by using MATLAB. The results of this measurement were compared to the results of calculation in the treatment planning. In the isodose curve on the axial direction, the dose distribution coincided with the high dose area, 0.76mm difference on 80%, rather than the low dose area, 1.29 mm difference on 40%. In this research, the fact that the polymer gel dosimeter and MRI can be applied for analyzing a small field in a 3 dimensional dosimetry was confirmed. Moreover, the feasibility of using these for the therapeutic radiation quality control was also confirmed.
중합체 겔 선량계를 제작하여 사이버나이프의 광자선의 소조사면에서의 3차원적 선량측정 실험을 수행하였다. 겔 선량계는 아크릴로 제작된 두경부 팬텀내에 설치하여 선량측정에 사용하였다. 두경부 팬텀의 영상은 사이버나이프의 두경부 환자 치료를 위해 사용하는 전산화단층촬영장치(CT)의 프로토콜을 사용하여 획득하였고 치료계획 시스템에 전달되었다. 겔 선량계에 치료계획에서 수립한 방사선을 조사하였으며, 24시간 경과 후에 선량분석을 위하여 3.0T 자기공명영상장치(MRI)로 영상을 획득하였다. 측정된 겔 팬텀의 선량분포는 MATLAB을 이용하여 분석하였고 측정된 결과를 치료계획에서의 계산값과 비교하였다. 선량분포는 축상(axial) 방향의 등선량곡선의 80% 고선량 영역에서 치료계획 선량곡선과 측정된 선량곡선과의 차이가 0.76 mm이었으며, 40% 저선량 영역에서는 차이가 1.29 mm로 저선량 영역보다 고 선량영역에서 잘 일치함을 알 수 있었다. 본 연구에서 중합체 겔 선량계와 자기공명영상을 이용하여 소조사면에 대해 3차원적으로 선량을 분석할 수 있을 뿐 아니라 치료방사선 정도관리에도 활용될 수 있는 가능성을 확인하였다.