Assessment of Set-up Accuracy in Tangential Breast Treatment Using Electronic Portal Imaging Device

EPID 영상을 이용한 유방암 접선조사의 정확성 평가

  • Lee, Byung-Koo (Department of Radiation Oncology, Korea University Anam Hospital) ;
  • Kang, Soo-Man (Department of Radiation Oncology, Kosin University Gospel Hospital)
  • 이병구 (고려대학교안암병원 방사선종양학과) ;
  • 강수만 (고신대학교 복음병원 방사선종양학과)
  • Received : 2012.05.10
  • Accepted : 2012.08.27
  • Published : 2012.09.28

Abstract

The aim of this study was to investigate the setup accuracy for tangential breast treatment patients using electronic portal image and 2-D reconstruction image Twenty two patients undergoing tangential breast treatment. To explore the setup accuracy, distances between chosen landmarks were taken as reference parameters. The difference between measured reference parameters on simulation films and electronic portal images (EPIs) was calculated as the setup error. A total of 22 simulation films and 110 EPIs were evaluated. In the tangential fields, the calculated reference parameters were the central lung distance (CLD), central soft-tissue distance (CSTD), and above lung distance (ALD), below lung distance (BLD). In the medial tangential field, the average difference values for these parameters were 1.0, -6.4, -2.1 and 2.0, respectively; and the ${\sigma}$ values were 1.5, 2.3, 4.1 and 1.1, respectively. In the lateral tangential field, the average difference values for these parameters were -1.5, -4.3, -2.7 and -1.3, respectively; and the ${\sigma}$ values were 3.3, 2.1, 2.9 and 2.5, respectively. CLD, CSTD, ALD and BLD in the tangential fields are easily identifiable and are helpful for detecting setup errors using EPIs in patients undergoing tangential breast radiotherapy treatment.

본 논문은 유방암 환자의 접선 조사 시, 치료 전 수행하는 electronic portal image와 치료 계획용 2-D reconstruction image를 비교하는 방법으로 자세의 정확성을 평가하고자 한다. 방사선 치료 중 접선조사(tangential breast treatment)만을 받는 22명의 유방암 환자를 대상으로 자세 정렬의 정확성을 확인 하였다. electronic portal image와 치료 계획용 2-D reconstruction image의 해부학적 기준 매개 변수를 비교하여 그 오차 도를 평가 하였다. 접선조사 환자의 44매 2-D reconstruction image와 110매의 EPID image 상의 비교 기준 매개 변수는 치료 조사면 중심부(field center)의 폐 길이, CLD(central lung distance), 치료 조사면 중심부의 연부조직 길이, CSTD(central soft tissue distance), 상부 총 폐 길이, ALD (above lung distance), 하부 총 폐 길이, BLD(below lung distance)이며, 내측 접선조사면(medial tangential field)에서 각 매개 변수의 오차 평균값은 1.0, -6.4, -2.1, 2.0, 각각의 표준편차(${\sigma}$)는 1.5, 2.3, 4.1, 1.1 이다. 외측 접선조사면(lateral tangential field)의 각 매개 변수 오차 평균값은 -1.5, -4.3, -2.7, -1.3 이며, 각각의 표준편차(${\sigma}$)는 3.3, 2.1, 2.9, 그리고 2.5로 나타났다. 접선조사 치료를 받는 유방암 환자의 EPID image 상에서 CLD, CSTD, ALD 그리고 BLD의 인식은 매우 용이하며 이를 근거로 자세 정렬 오차를 판단하는 것이 시간과 숙련도의 단축을 이끌어 낼 수 있다고 사료된다.

Keywords

References

  1. 보건복지부: 2010보건복지통계연보. 제56호 p. 60.
  2. 한국유방암학회: 유방암 등록사업을 프로그램을 이용한 2004년 전국적인 한국인 유방암 자료 분석, J Breast Cancer. 9(2): 151-161, 2006 https://doi.org/10.4048/jbc.2006.9.2.151
  3. 지영훈, 김미숙, 류성렬 등: 방사선종양학과 전국 통계(2006년). 대한 방사선 종양학회지, 26, 131-133. 2008
  4. Clarke M, Collins R, Darby S, Davies C, Elphinstone P, Evans E, Godwin J, Gray R, Hicks C, James S, MacKinnon E, McGale P, McHugh T, Peto R, Taylor C, Wang Y. Early Breast Cancer Trialists' Collaborative Group (EBCTCG). Effects of radiotherapy and of differences in the extent of surgery for early breast cancer on local recurrence and 15-year survival: an overview of the randomised trials. Lancet, 366, 2087- 2106, 2005 https://doi.org/10.1016/S0140-6736(05)67887-7
  5. Vinh-Hung V, Verschraegen C. Breast-conserving surgery with or without radiotherapy: Pooled-analysis for risks of ipsilateral breast tumor recurrence and mortality. J Natl Cancer Inst, 96, 115-121, 2004 https://doi.org/10.1093/jnci/djh013
  6. Overgaard M, Hansen PS, Overgaard J, Rose C, AnderssonM, et al. Postoperative radiotherapy in high-risk premenopausal women with breast cancer who receive adjuvant chemotherapy: Danish Breast Cancer Cooperative Group 82b Trial. N Engl J Med, 337, 949-955, 1997 https://doi.org/10.1056/NEJM199710023371401
  7. Landau D, Adams EJ, Webb S, Ross G. Cardiac avoidance in breast radiotherapy: a comparison of simple shielding techniques with intensity- modulated radiotherapy. Radiotherapy and Oncology, 60, 247-55. 2001
  8. Hardenbergh PH, Bentel GC, Prosnitz LR, Marks LB. Postmastectomy radiotherapy: toxicities and techniques to reduce them. Semin Radiat Oncol, 9, 259-268, 1999
  9. Coen W. Hurkmans, Peter Remeijer, Joos V. Lebesque, Ben J. Mijnheer. Set-up verification using portal imaging; review of current clinical practice. Radiotherapy and Oncology, 58, 105- 120, 2001 https://doi.org/10.1016/S0167-8140(00)00260-7
  10. Lujan AE, Larsen EW, Balter JM, et al: A method for incorporating organ motion due to breathing into 3D dose calculations. Med Phys, 26, 715-720, 1999 https://doi.org/10.1118/1.598577
  11. Fiorino C, Reni M, Bolognesi A, et al: Intra-and interobserver variability in contouring prostate and seminal vesicles: Implications for conformal treatment planning. Radiotherapy and Oncology, 47, 285-292, 1998 https://doi.org/10.1016/S0167-8140(98)00021-8
  12. Dutreix A. When and how can we improve precision in radiotherapy? Radiotherapy and Oncology, 2, 275-292, 1984 https://doi.org/10.1016/S0167-8140(84)80070-5
  13. Luchka K, Shalev S. Pelvic irradiation of the obese patient: a treatment strategy involving megavoltage simulation and intratreatment setup corrections. Med Phys, 23, 1897-1902, 1996 https://doi.org/10.1118/1.597752
  14. A. Lirette, J. Pouliot, M. Aubin, M. Larochelle. The role of electronic portal imaging in tangential breast irradiation:a prospective study. Radiotherapy and Oncology, 37, 241-245, 1995 https://doi.org/10.1016/0167-8140(95)01653-8
  15. Van den Heuvel F, De Neve W, Verellen D, et al. Clinical implementation of an objective computer- aided protocol for intervention in intra- treatment correction using electronic portal imaging. Radiotherapy and Oncology, 35, 232- 239, 1995 https://doi.org/10.1016/0167-8140(95)01569-3
  16. van Tienhoven, G., Lanson, J.H., Crabeels, D., Heukelom, S. and Mijnheer, B.J. Accuracy in tangential breast treatment set up: a portal imaging study. Radiotherapy and Oncology, 22, 317-322, 1991 https://doi.org/10.1016/0167-8140(91)90171-C
  17. Herman MG, Abrams RA, Mayer RR. Clinical use of on-line portal imaging for daily patient treatment verification. Int J Radiat Oncol Biol Phys, 28, 1017-1023, 1994 https://doi.org/10.1016/0360-3016(94)90123-6
  18. De Neve W, Van den Heuvel F, De Beukeleer M, et al. Routine clinical on-line portal imaging followed by immediate field adjustment using a tele-controlled patient couch. Radiotherapy and Oncology, 24, 45-54, 1992 https://doi.org/10.1016/0167-8140(92)90353-V