Evaluation of IH-1000 for Automated ABO-Rh Typing and Irregular Antibody Screening

ABO 및 RhD 혈액형 검사와 비예기항체 선별검사를 위한 자동화장비 IH-1000의 평가

  • Park, Youngchun (Department of Laboratory Medicine, College of Medicine, Chungnam National University) ;
  • Lim, Jinsook (Department of Laboratory Medicine, College of Medicine, Chungnam National University) ;
  • Ko, Younghuyn (Department of Laboratory Medicine, College of Medicine, Chungnam National University) ;
  • Kwon, Kyechul (Department of Laboratory Medicine, College of Medicine, Chungnam National University) ;
  • Koo, Sunhoe (Department of Laboratory Medicine, College of Medicine, Chungnam National University) ;
  • Kim, Jimyung (Department of Laboratory Medicine, College of Medicine, Chungnam National University)
  • 박영춘 (충남대학교 의학전문대학원 진단검사의학교실) ;
  • 임진숙 (충남대학교 의학전문대학원 진단검사의학교실) ;
  • 고영현 (충남대학교 의학전문대학원 진단검사의학교실) ;
  • 권계철 (충남대학교 의학전문대학원 진단검사의학교실) ;
  • 구선회 (충남대학교 의학전문대학원 진단검사의학교실) ;
  • 김지명 (충남대학교 의학전문대학원 진단검사의학교실)
  • Received : 2012.07.20
  • Accepted : 2012.08.21
  • Published : 2012.08.31

Abstract

Background: Despite modern advances in laboratory automated medicine, work-process in the blood bank is still handled manually. Several automated immunohematological instruments have been developed and are available in the market. The IH-1000 (Bio-Rad Laboratories, Hercules, CA, USA), a fully automated instrument for immunohematology, was recently introduced. In this study, we evaluated the performance of the IH-1000 for ABO/Rh typing and irregular antibody screening. Methods: In October 2011, a total of 373 blood samples for ABO/Rh typing and 303 cases for unexpected antibody screening were collected. The IH-1000 was compared to the manual tube and slide methods for ABO/Rh typing and to the microcolumn agglutination method (DiaMed-ID system) for antibody screening. Results: For ABO/Rh typing, concordance rate was 100%. For unexpected antibody screening, positive results for both column agglutination and IH-1000 were observed in 10 cases (four cases of anti-E and c, three of anti-E, one of anti-D, one of anti-M, and one of anti-Xg) and negative results for both were observed in 289 cases. The concordance rate between IH-1000 and column agglutination was 98.7%. Sensitivity and specificity were 90.9% and 99.3%, respectively. Conclusion: The automated IH-1000 showed good correlation with the manual tube and slide methods and the microcolumn agglutination method for ABO-RhD typing and irregular antibody screening. The IH-1000 can be used for routine pre-transfusion testing in the blood bank.

배경: 대부분의 임상검사가 자동화되는 추세이나 혈액은행 검사는 여전히 수작업에 의존하고 있다. 여러 자동화장비가 개발되어 활용되고 있으며 최근 면역혈액검사를 위한 자동화장비인 IH-1000 (Bio-Rad Laboratories, Hercules, CA, USA)이 소개되었다. 이번 연구에서는 IH-1000 장비의 ABO 및 Rh 혈액형 검사와 비예기항체 선별검사에 대한 성능을 평가하려고 한다. 방법: 2011년 10월에 의뢰된 373건의 ABO 및 RhD 혈액형 검사와 303건의 비예기항체 선별검사를 대상으로 하였다. ABO 및 RhD 혈액형 검사는 슬라이드법과 시험관법을 사용하였고 비예기항체검사는 미세원주응집법(DiaMed-ID system)을 사용하여 IH-1000의 결과와 비교하였다. 결과: ABO 및 RhD 혈액형 검사에서는 두 방법 간 100%의 높은 일치율을 보였고 비예기항체 선별검사에서는 두 방법에서 모두 양성을 보인 검체가 10건(anti-E+c 4건, anti-E 3건, anti-D 1건, anti-M 1건, anti-Xg 1건)이었고 모두 음성을 보인 검체는 289건으로 98.7%의 일치율을 보였다. IH-1000의 민감도 및 특이도는 각각 90.9%였다. 결론: IH-1000은 기존에 사용하던 수기법과 비교하여 ABO 및 Rh 혈액형 검사와 비예기항체 선별검사에서 높은 일치도를 보였다. 따라서 혈액은행에서의 수혈 전 일상검사에 자동화 장비인 IH-1000을 사용할 수 있을 것으로 생각된다.