DOI QR코드

DOI QR Code

B1+ Comparison for Different RF Coils at 7 T MRI

7 T MRI에서의 다양한 RF 코일에 대한 B1+ 비교

  • Kim, Hong-Joon (Department of Electrical Engineering, IT College, Kyungpook National University) ;
  • Yoo, Hyoung-Suk (Department of Biomedical Engineering, School of Electrical Engineering, University of Ulsan)
  • 김홍준 (경북대학교 전기공학과) ;
  • 유형석 (울산대학교 전기공학부 의공학과)
  • Received : 2012.06.12
  • Accepted : 2012.07.03
  • Published : 2012.08.31

Abstract

This paper demonstrates the use of the convex optimization to localize the transverse magnetic $B_1^+$ field in regions of interest for recently proposed multi-sectioned alternating impedance coils and the traditional transmission line coil. An approach based on different axial slices to identical RF coils except upper stripline structure is investigated. Electromagnetic simulation results are compared for RF coils and discussed in detail at 7.0 T.

정 자장($B_0$)의 세기가 7 T(Tesla) 고자기장 MRI(Magnetic Resonance Imaging) 시스템은 정 자장의 세기가 1.5 T 또는 3 T MRI 시스템에 비하여 인가된 RF(Radio Frequency) 필드의 높은 불균질성을 보여준다. 이러한 문제점를 극복하기 위한 한 방법으로써, 관심 영역에서의 RF 자장이 균일하도록 RF 코일의 각 요소로 들어가는 최적화된 전류의 크기와 위상 값을 얻기 위해서 convex 최적화 방법이 사용된다. 이러한 방법을 7 T MRI 시스템에 다양한 RF 코일에 적용하여 각각의 $B_1^+$ 필드 값을 비교하여 그 성능을 파악하였다.

Keywords

References

  1. J. T. Vaughan, M. Garwood, C. M. Collins, W. Liu, L. Delabarre, G. Adriany, P. Andersen, H. Merkle, R. Goebel, M. B. Smith, and K. Ugurbil, "7T vs 4- T: RF power, homogeneity, and signal-to-noise comparison in head images", Magn Reson Med, vol. 46, no. 1, pp. 24-30, Jul. 2001. https://doi.org/10.1002/mrm.1156
  2. J. T. Vaughan, L. DelaBarre, C. Snyder, J. Tian, C. Akgun, D. Shrivastava, W. Liu, C. Olson, G. Adriany, J. Strupp, P. Andersen, A. Gopinath, P. van de Moortele, M. Garwood, and K. Ugurbil, "9.4T human MRI: Preliminary results", Magn Reson Med, vol. 56, no. 6, pp. 1274-1282, Dec. 2006. https://doi.org/10.1002/mrm.21073
  3. J. Vaughan, "RF Coil for Imaging System", U.S. patent 6,633:161, 2003.
  4. C. Olson, H. Yoo, L. Delabarre, J. T. Vaughan, and A. Gopinath, "RF B1 field localizations through convex optimization", Microwave and Optical Technology Letters, vol. 54, pp. 31-37, Jan. 2011.
  5. 김홍준, 손혁우, 조영기, 유형석, "고자장 MRI에 서의 영상 영역에 대한 $B^{+}_{1}$ 균질성", 한국전자파학회논문지, 23(1), pp, 96-100, 2012년 1월.