Changes in External Radiation Dose Rate for PET-CT Test Patients

PET-CT 검사 환자의 외부 방사선량률 변화

  • Kim, Su-Jin (Department of Radiologic Technology, Daegu Health College) ;
  • Han, Eun-Ok (Department of Radiologic Technology, Daegu Health College)
  • Received : 2012.04.30
  • Accepted : 2012.05.21
  • Published : 2012.06.30

Abstract

This paper analyzes changes in the external radiation dose rate of PET-CT test patients as a part of providing basic materials for reduction of radiation exposure to PET-CT test patients. In theory the measurement of external radiation dose rate of PET-CT test patients shows that the further the distance from the patient injected with radioactive pharmaceutical and a longer time elapsement from the injection leads to a smaller amount of radiation. Particularly, the amount of radiation marked the highest in the chest was at 4.17 minutes immediately after the intravenous injection and in the head after 77.47 minutes after urination in advance to the PET-CT test. As in the generalized information, it is desired to keep distance between the patient and caretakers or professionals to reduce the amount of radiation exposure from PET-CT test patients and to resume contact the patient after the time when the radiation has reduced. If contact is unavoidable, it is desired to keep at least 200cm from the patient. In addition, the amount of radiation reached the highest in the chest at first and then in the head from 77 minutes after injection. Accordingly, it would be helpful in achieving the optimization if contact is made based on the patient's physical characteristics. This study is significant as it measures changes in radiation the dose rate by; distance from the PET-CT test patient, time elapsed, and specific parts of body. Further studies based on the findings in this paper are required to analyze changes in radiation dose rate in accordance with individual characteristics unique to PET-CT patients and to utilize the results to reduce the amount of radiation patient, caretakers and professions are exposed.

PET-CT 검사 환자의 피폭선량 감소를 위한 기초자료 제공의 일환으로 PET-CT 검사 환자의 방사선량률의 변화를 분석하고자 하였다. PET-CT 검사 환자의 방사선량률을 측정한 결과 이론과 같이 방사성의약품이 투여된 환자로부터 거리가 멀수록, 시간이 지날수록 방사선량률은 감소되는 것을 볼 수 있었다. 특히 신체부위에 따라서는 방사성의약품 정맥 주사 즉시인 약 4.17분에서는 흉부, PET-CT 검사 전 배뇨 후인 약 77.47분 이후부터는 두부가 가장 높게 나타났다. 일반화되어 있는 정보와 같이 PET-CT 검사 환자로부터 받는 방사선 피폭량을 감소시키기 위해서는 보호자나 방사선작업종사자가 환자로부터 거리를 멀리하거나 방사능이 감소된 이후의 시간부터 접촉하는 것이 바람직하다. 불가피한 접촉이 필요하다면 가능한 거리는 200 cm이상을 확보하는 것이 바람직하다. 또한 초기에는 흉부, 방사성의약품 투여 후 약 77분 이후부터는 두부에 방사선량률이 높기 때문에 환자 신체적 특징을 고려한 접촉도 함께 이루어진다면 최적화 달성에 도움이 될 것이라고 보여 진다. 본 연구에서 도출된 PET-CT 검사 환자의 거리, 시간, 신체부위에 따른 방사선량률 변화를 알 수 있다는 점에서 연구에 의의가 있다고 본다. 향후 연구에서는 본 연구에서 도출된 결과를 바탕으로 환자 개인특성에 따른 방사선량률의 변화 차이를 분석하여 환자, 보호자, 종사자의 피폭선량 감소에 활용할 수 있도록 지속적인 연구가 수행되는 것이 필요하다고 본다.

Keywords

References

  1. 대한핵의학회. http://www.ksnm.or.kr/. 2010.
  2. 이화수. 18F-FDG를 이용한 PET/CT 검사 시 피폭선량과 혈구 변화. 부산카톨릭대학교 생명과학대학원 학위논문. 2011.
  3. 남소라, 손혜경, 이상훈, 이창래, 조효민, 김희중. 전신 PET/CT 영상 획득 프로토콜을 이용한 유효선량 평가. 한국의학물리학회지:의학물리 2006;17(3):173-178.
  4. 임창선, 김세헌. 핵의학과에서 방사선 피폭관리 실태에 대한 조사 연구. 한국산학기술학회논문지. 2009;10(7):1760-1765.
  5. 정진욱, 한은옥. PET-CT의 방사선안전관리 가이드라인 개발: 행위와 환경을 중심으로. 방사선방어학회지. 2011;36(3):140-147.
  6. 한양대학교 산학협력단. 의료방사선 안전에 관한 국제방사선방호위원회(ICRP) 동향연구. 과학기술부. 2006.
  7. 한은옥. 의료기관 방사선종사자의 방사선장해 방어행위 모형. 이화여자대학교 대학원 학위논문. 2009.
  8. 이재기. ICRP 105 의료에서의 방사선방호 역자서문. 2010.
  9. 곽병준. PET-CT 검사에서 공간선량률 측정에 관한 연구. 부산카톨릭대학교 보건과학대학원 학위논문. 2009.
  10. 박명환. 전신 PET/CT 검사에서 공간선량률 측정. 방사선기술과학. 2006;29(4):257-260.
  11. 손혜경, 이상훈, 남소라, 김희중. 전신 PET/CT 영상 획득 시 투과 스캔에서의 방사선 선량. 한국의학물리학회지:의학물리 2006;17(2):89-95.
  12. Almeida P, Bendriem B, de Dreuille O, Peltier A, Perrot C, Brulon V. Dosimetry of transmission measurements in nuclear medicine: a study using anthropomorphic phantoms and thermoluminescent dosimeters. Eur. J. Nucl. Med. 1998;25(10):1435-1442. https://doi.org/10.1007/s002590050320
  13. Brix G, Lechel U, Glatting G, Ziegler SI, Munzing W, Muller SP, Beyer T. Radiation exposure of patients undergoing whole-body dual-modality 18F-FDG PET/CT examinations. J. Nucl. Med. 2005;46(4):608-613.
  14. Wu TH, Huang YH, Lee JJ, Wang SY, Wang SC, Su CT, Chen LK, Chu TC. Radiation exposure during transmission measurements: comparison between CT- and germanium-based techniques with a current PET scanner. Eur. J. Nucl. Med. Mol. Imaging. 2003;31(1):38-43.
  15. 곽병준, 지태정, 민병인. F-18 FDG를 이용한 핵의학 검사에서 주변 선량의 안전성 평가. 한국안전학회지. 2009;24(6):157-162.
  16. 김봉수, 표성재, 조용귀, 신채호, 조진우, 김창호. PET/CT 검사 환자의 피폭선량 경감을 위한 연구. 핵의학기술학회지. 2009;13(3):10-16.
  17. 권오진, 정서희, 백승찬, 김경호. 근무형태 및 피폭선량에 따른 PET/CT실의 운영 효율 성 평가. 핵의학기술학회지. 2009;13(1):93-97.
  18. 이병일. 핵의학 영상장비 PET/CT의 정도관리와 성능평가. 대한핵의학회지. 2008;42(2):137-144.
  19. Benatar NA, Cronin BF, O'Doherty MJ. Radiation dose rates from patients undergoing PET: implications for technologists and waiting areas. Eur. J. Nucl. Med. 2000;27(5):583-589. https://doi.org/10.1007/s002590050546
  20. White S, Binns D, Johnston V, Fawcett M, Greer B, Ciavarella F, Hicks R. Occupational exposure in nuclear medicine and PET. Clin. Positron Imaging. 2000;3(3):127-129. https://doi.org/10.1016/S1095-0397(00)00044-3
  21. 안샤론, 박훈희, 박민수, 오신현, 이승재, 임한상, 김재삼, 이창호. PET/CT 영상에서 조영제를 이용한 CT 영상의 보정(Correction)에 따른 표준화섭취계수 (SUV)의 영향. 핵의학기술학회지. 2009;13(1):77-81.
  22. 박훈희, 김정열, 이승재, 박민수, 남궁혁, 임한상, 오기백, 김재삼, 이창호, 진계환. PET/CT에서 표준섭취계수(SUV)의 신뢰성 확보를 위한 초기연구. 핵의학기술학회지. 2009;13(3):31-42.
  23. 신규설, 동경래. PET-CT에서 CT의 관전압 및 관전류에 따른 SUV값의 변화. 방사선기술과학. 2007;30(4):373-379.
  24. 남소라, 손혜경, 임한상, 박훈희, 조효민, 이창래, 김희중. 전신 PET/CT 영상에서 조영제 영향의 보정 유.무에 따른 SUV 평가. 한국의학물리학회지:의학물리. 2006;17(4): 246-251.
  25. 정영진, 강도영. 진단용 고선량 CT를 이용한 18FFDGPET/CT 촬영시 정맥 조영제가 SUV 값에 미치 는 영향. 대한핵의학회지. 2006;40(3):169-176.
  26. Mawlawi O, Erasmus JJ, Munden RF, Pan T, Knight AE, Macapinlac HA, Podoloff DA, Chasen M. Quantifying the effect of IV contrast media on integrated PET/CT: Clinical Evaluation. Am. J. Roentgenol. 2006;186:308-319. https://doi.org/10.2214/AJR.04.1740
  27. Yau YY, Chan WS, Tam YM, Vernon P, Wong S. Coel M. Chu SK. Application of intravenous contrast in PET/CT: does it really introduce significant attenuation correction error?. J. Nucl. Med. 2005;46(2):283-291.
  28. Beyer T, Antoch G, Bockisch A, Stattaus J. Optimized intravenous contrast administration for diagnostic whole-body 18F-FDG PET/CT. J. Nucl. Med. 2005;46(3):429-435.
  29. 서울대학교 의과대학 편. 내분비학. 서울대학교출판문화원. 2005:434.

Cited by

  1. Measurement and Estimation for the Clearance of Radioactive Waste Contaminated with Radioisotopes for Medical Application vol.25, pp.1, 2014, https://doi.org/10.14316/pmp.2014.25.1.8
  2. Solution to Decrease Spatial Dose Rate in Laboratory of Nuclear Medicine through System Improvement vol.20, pp.1, 2014, https://doi.org/10.14371/QIH.2014.20.1.60
  3. CT 촬영 조건에 따른 PET 영상의 변화 vol.13, pp.3, 2019, https://doi.org/10.7742/jksr.2019.13.3.473