Evaluation of SharpIR Reconstruction Method in PET/CT
PET/CT 검사에서 SharpIR 재구성 방법의 평가
- Kim, Jung-Yul (Department of Nuclear Medicine, Yonsei University Health System) ;
- Kang, Chun-Koo (Department of Nuclear Medicine, Yonsei University Health System) ;
- Park, Hoon-Hee (Department of Radiology, Shingu University) ;
- Lim, Han-Sang (Department of Nuclear Medicine, Yonsei University Health System) ;
- Lee, Chang-Ho (Department of Nuclear Medicine, Yonsei University Health System)
- 김정열 (연세의료원 세브란스병원 핵의학과) ;
- 강천구 (연세의료원 세브란스병원 핵의학과) ;
- 박훈희 (신구대학교 방사선학과) ;
- 임한상 (연세의료원 세브란스병원 핵의학과) ;
- 이창호 (연세의료원 세브란스병원 핵의학과)
- Received : 2012.02.29
- Accepted : 2012.04.06
- Published : 2012.04.28
Abstract
Purpose : In conventional PET image reconstruction, iterative reconstruction methods such as OSEM (Ordered Subsets Expectation Maximization) have now generally replaced traditional analytic methods such as filtered back-projection. This includes improvements in components of the system model geometry, fully 3D scatter and low noise randoms estimates. SharpIR algorithm is to improve PET image contrast to noise by incorporating information about the PET detector response into the 3D iterative reconstruction algorithm. The aim of this study is evaluation of SharpIR reconstruction method in PET/CT. Materials and Methods: For the measurement of detector response for the spatial resolution, a capillary tube was filled with FDG and scanned at varying distances from the iso-center (5, 10, 15, 20 cm). To measure image quality for contrast recovery, the NEMA IEC body phantom (Data Spectrum Corporation, Hillsborough, NC) with diameters of 1, 13, 17 and 22 for simulating hot and 28 and 37 mm for simulating cold lesions. A solution of 5.4 kBq/mL of
서론: 종래의 PET 영상 재구성에 있어서 FBP 등에 비해 3차원 반복 재구성 방법이 일반적으로 대체하고 있으며, 이것은 검출기 기하학적 특성과 완벽한 3차원 산란 평가 및 저잡음 randoms 평가 등의 더 진보된 재구성 알고리즘을 제공하고 활용되고 있다. 최근에 SharpIR알고리즘은 3차원 반복 재구성 알고리즘으로 PET 검출기 응답 정보를 통합하여 PET 영상의 잡음을 효과적으로 감소시켜 대조도를 향상 시키기 위한 것으로 알려지고 있다. 본 연구에서는 새로운 반복 시스템 모델인 SharpIR에 대한 성능 평가와 임상에서의 적용 가능성에 대해 알아보고자 한다. 실험재료 및 방법: 검출기 응답에 대한 분해능을 측정하기 위해 유리관(내경 1.1 mm, 두께 0.2 mm)에