복부일반촬영시 선원과 검출기간의 거리변화에 따른 영상 화질 및 피폭선량에 관한 연구
A Study on Patients Dose and Image Quality according to Source to Image receptor Distance in Abdomen Radiography: comparison of ESD measured and DRLs in other countries
- 장지성 (서울아산병원 영상의학팀) ;
- 최원근 (서울아산병원 영상의학팀) ;
- 정재연 (서울아산병원 영상의학팀) ;
- 이관섭 (서울아산병원 영상의학팀) ;
- 하동윤 (서울아산병원 영상의학팀)
- Jang, Ji-Sung (Department of Radiology, Asan Medical Center) ;
- Choi, Weon-Keun (Department of Radiology, Asan Medical Center) ;
- Jung, Jae-Yon (Department of Radiology, Asan Medical Center) ;
- Lee, Kwan-Sub (Department of Radiology, Asan Medical Center) ;
- Ha, Dong-Yoon (Department of Radiology, Asan Medical Center)
- 투고 : 2012.09.14
- 심사 : 2012.10.26
- 발행 : 2012.10.31
초록
Purpose : The purpose of this study was to reduce Entrance Surface Dose and maintain image quality by changing Source to Image receptor Distance. And we'd like to compare ESD on this study to DRLs in other contries. Materials and Methods : We used indirect DR system(Definium 8000, General Electric, USA)and phantom(ART-200X, Flukebiomedical, USA),glass dosimeters(GD-352M, Asahi Techno Glass, Japan)for this study. The imagies were obtained throuh 80kVp fixed, and different tube currents using AEC mode in