Investigation of Response of Several Neutron Surveymeters by a DT Neutron Generator

DT 중성자 발생기에 의한 중성자 검출기 반응도 조사

  • Received : 2012.01.30
  • Accepted : 2012.02.23
  • Published : 2012.03.30

Abstract

Several neutron measuring devices were tested under the neutron fields characterized with two distinct kinds of thermal and fast neutron spectrum. These neutron fields were constructed by the mixing of both thermal neutron fields and fast neutron fields. The thermal neutron field was constructed using by a graphite pile with eight AmBe neutron sources. The fast neutron field of 14 MeV was made by a DT neutron generator. In order to change the fraction of fast neutron fluence rate in each neutron fields, a neutron generator was placed in the thermal neutron field at 50 cm and 150 cm from the reference position. The polyethylene neutron collimator was used to make moderated 14 MeV neutron field. These neutron spectra were measured by using a Bonner sphere system with an LiI scintillator, and dosimetric quantities delivered to neutron surveymeters were determined from these measurement results.

국내 교정기관 또는 표준기관은 중성자 검출기의 교정을 위해 비감속 및 중수감속 $^{252}Cf$ 선원과 $^{241}AmBe$ 선원을 사용하고 있다. 이런 선원들로 교정된 중성자 검출기를 이용하여 입자가속기와 같이 속중성자가 다량 존재하는 시설을 선량평가할 때, 그 정확도가 떨어지게 된다. 그 이유는, 대부분의 중성자 검출기는 열중성자에 민감하게 반응하므로 수 MeV 이상의 에너지를 가지는 속중성자장에 대한 선량당량 반응도는 부정확하다. 또한 높은 에너지의 중성자는 열중성자보다 선량기여정도가 훨씬 크기 때문이다. 이와 같은 이유로, 기존의 교정용 기준 중성자장이 아닌 수 MeV 이상의 속중성자가 존재하는 중성자장에서도 검출기를 교정할 필요가 있다. DT 중성자 발생기, 흑연집합체 그리고 폴리에틸렌 중성자 집속체를 사용하여 속중성자의 선속분율이 서로 다른 중성자장을 제작하였고, 이 중성자장에서 중성자 검출기의 선량당량 반응도를 측정하였다. 시험결과에 의하면, 속중성자 선속분율과 중성자 검출기의 종류에 따라 중성자 검출기의 반응도는 많은 차이를 보였다. 이러한 반응도 차이는 선량당량의 과대 및 과소평가를 의미하므로, 검출기가 사용되는 시설환경과 유사한 중성자장에서 반응도 교정이 필요함을 확인하였다.

Keywords

References

  1. KAERI. Technology development for radiation dose measurement and evaluation. KAERI/RR-2801. 2007.
  2. Knoll GF, Radiation detection and measurement. 3rd ed.. New York; John Wiley & Sons, Inc.. 2000:506-510.
  3. Alvera AV and Thomas DJ. Neutron field spectrometry in mixed fields: multisphere spectrometers. Radiat. Prot. Dosim. 107(1-3):37-72; 2003. https://doi.org/10.1093/oxfordjournals.rpd.a006388
  4. ISO. Reference neutron radiation-part 1: characteristics and methods of production. ISO 8529-1: 1st ed.. 2001.
  5. Lisken H. and Paulsen A., Neutron production cross section and energies for the reactions $T(p,n)^3He,D(d,n)^3He$ and $T(d,n)^4He$. Atom. Data Nucl. Data 11:569-619; 1973. https://doi.org/10.1016/S0092-640X(73)80081-6
  6. EADS SODERN. GENIE 16C/D user manual, ver. July 2008. Eads Sodern Inc.. 2008.
  7. Thomas DJ and Kolkowski P. Thermal fluence and dose equivalent standards at NPL. DQL RN008. National Physics Laboratory UK. 2005.
  8. Lacoste V., Gressier V., Muller H. Lebreton L. Characterization of the IRSN graphite moderated americium-beryllium neutron field. Radiat. Prot. Dosim. 110(1-4):135-139; 2004. https://doi.org/10.1093/rpd/nch188
  9. Kim BH, Jun SM, Kim JS, Lim KS, Kim JL. Construction of thermal neutron calibration fields using a graphite pile and americium-beryllium neutron sources at KAERI. Nucl. Technol. 168(2): 349-353; 2009. https://doi.org/10.13182/NT09-A9207
  10. Pelowitz DB. MCNPX user's manual, version 2.5.0. LA-CP-05-0369. 2005.
  11. ICRU. Conversion coefficient for use in radiological protection against external radiation. ICRU Report 57. 1998.