Abstract
The Laser-Induced Breakdown Spectroscopy (LIBS) has great advantages as an analytical technique, namely real-time analysis without sample preparation, ideal for mobile chemical sensor for space exploration. The LIBS plasma characteristics are strongly dependent on the surrounding pressure. In this study, seven types of target (C, Ti, Ni, Cu, Sn, Al, Zn) were investigated for their elemental lifetime. The target was located in vacuum chamber which has the pressure range of 760 to $10^{-5}$ torr. As the pressure is decreased, the elemental lifetimes of carbon and titanium declined, while all other targets showed increased lifetimes until reaching 1 torr and declined with continued pressure decrease. The boiling point and electronegativity amongst the physicochemical properties of the samples are used to explain this peculiarity.
레이저 조사 시 발생되는 플라즈마를 성분 분석에 이용하는 Laser-Induced Breakdown Spectroscopy(LIBS)는 우주 자원 탐사에 적합한 실시간 성분 분석 기술이다. LIBS 플라즈마의 특성은 주위 압력의 영향을 크게 받는다. 본 연구에서는 다양한 물성치를 갖는 일곱 가지 원소(C, Ti, Ni, Cu, Sn, Al, Zn)의 지속시간(lifetime)이 760 - $10^{-5}$ torr의 압력 범위에서 분석되었다. 압력이 낮아짐에 따라 탄소와 티타늄의 lifetime은 감소하였고, 그 밖의 원소들은 1 torr의 압력에서 가장 오랜 시간동안 검출되었다. 원소별 lifetime 결과를 통하여 낮은 압력 하에서 플라즈마의 특성과 원소별 끓는점 및 전기음성도의 관계를 규명하였다.