Abstract
We studied the optimal location where the radiation dose of radiological technologists is minimally measured. The measured scatter dose has been compared with the distance inverse square law. We measured the primary X-ray with different tube conditions (60, 70, 81 and 90 kVp) and distances (60, 120 and 180 cm). The scatter ray has been measured with various locations (42.5, 52.4 and 62.4 cm for front and back side, 0 to 60 cm with 10 cm interval for left and right side). The results of this study showed that the dose of primary X-ray was attenuated to 20.52 (27.20%), 28.58 (25.20%), 38.82 (26.32%) and 48.20 mR (26.27%) for each tube voltages at 120 cm. In addition, The dose were 7.06 (8.91%), 9.90 (8.73%), 13.64 (9.25%) and 16.60 mR (9.05%) at 180 cm. As for the scatter in front and back side, the attenuated dose were 0.15 mR (23.09%) and 0.15 mR (22.08%) at 120 cm, and 0.07 mR (10.43%) and 0.06 mR (8.83%) at 180 cm. Scatter was decreased in third quadrant. Therefore, it is recommended that radiological technologists should keep long distance from the object.
거리 역자승 법칙에 관한 엑스선 감약의 정도를 파악하고 이를 산란선 발생에서 적용하여 방사선사의 피폭을 저감할 수 있는 공간을 찾도록 알아보고자 하였다. 관전류량 10 mAs, 관전압 60 kVp, 70 kVp, 81 kVp, 90 kVp, 각각의 거리 60 cm, 120 cm, 180 cm에서 1차 선량을 측정하고, 산란선은 관전류량 20 mAs, 관전압 70 kVp, phantom의 중심부로부터 전면과 후면으로 42.5 cm, 52.5 cm, 62.5 cm 떨어진 곳에서 음 양극(좌우측)으로 각각 10 cm씩 60 cm까지 측정하였고, 거리역자승법칙과 비교하기 위해 전 후방 각각 42.5 cm, 85 cm, 127.5 cm에서 산란선을 측정하였다. 1차선은 거리가 2배에서는 20.52 mR(27.20%), 28.58 mR(25.20%), 38.82 mR(26.32%), 48.20 mR(26.27%)로 감약되고 거리가 3배에서는 7.06 mR(8.91%), 9.90 mR(8.73%), 13.64 mR(9.25%), 16.60 mR(9.05%)로 감약되는 것을 알 수 있었고, 산란선은 전 후방 각각 거리가 2배에서는 0.15 mR(23.09%), 0.15 mR(22.08%) 3배에서는 0.07 mR(10.43%), 0.06 mR(8.83%)로 감약되었다. 산란선의 발생량이 평균적으로 3사분면이 적게 발생하기에 환자를 붙잡고 촬영할 시에는 3사분면의 피사체에서 가능하면 거리를 두고 환자를 잡는 것이 피폭선량을 줄일 수 있다.