Evaluation of Dosimetry and Image of Very Low Dose CT Attenuation Correction for Pediatric PET/CT: Phantom Study
팬텀을 이용한 소아 PET/CT 검사 시 감쇄보정 CT 선량과 영상 평가
- Bahn, Young-Kag (Department of Nuclear Medicine, Severance Hospital, Yonsei University Health System) ;
- Kim, Jung-Yul (Department of Nuclear Medicine, Severance Hospital, Yonsei University Health System) ;
- Park, Hoon-Hee (Department of Radiology, Shingu University) ;
- Kang, Chun-Goo (Department of Nuclear Medicine, Severance Hospital, Yonsei University Health System) ;
- Lim, Han-Sang (Department of Nuclear Medicine, Severance Hospital, Yonsei University Health System) ;
- Lee, Chang-Ho (Department of Nuclear Medicine, Severance Hospital, Yonsei University Health System)
- 반영각 (연세의료원 세브란스병원 핵의학과) ;
- 김정열 (연세의료원 세브란스병원 핵의학과) ;
- 박훈희 (신구대학교 방사선과) ;
- 강천구 (연세의료원 세브란스병원 핵의학과) ;
- 임한상 (연세의료원 세브란스병원 핵의학과) ;
- 이창호 (연세의료원 세브란스병원 핵의학과)
- Received : 2011.06.29
- Accepted : 2011.07.05
- Published : 2011.11.05
Abstract
Purpose: To evaluate the dosimetry and image of very low does CT attenuation correction for phantom using pediatric PET/CT. Materials and methods: three PET / CT scanners (Discovery STe, BiographTruepoint 40, Discovery 600) as a child-size acrylic phantom and ion chamber dosimeter (Unfous Xi CT, Sweden) using a CT image acquisition parameters (10, 20, 40, 80, 100, 160 mA; 80, 100, 120, 140 kVp) by varying the depth dose and evaluate
최근 PET/CT 검사는 성인 암환자뿐만 아니라 소아 암환자에게도 많이 적용하고 있다. 그러나 소아 환자의 경우 성인에 비하여 방사성 감수성이 높아 피폭선량의 관리를 필요로 하고 있다. 그러므로 본 저자는 팬텀을 이용하여 소아 PET/CT 검사 시 감쇄보정 CT 선량과 영상을 평가하였다. 3대의 PET/CT 장비 (Discovery STe, BiographTruepoint 40, Discovery 600)로 소아 사이즈의 아크릴 팬텀과 이온 챔버 선량계 (Unfous Xi CT, Sweden)를 이용하여 CT 영상획득 조건 (10, 20, 40, 80, 100, 160 mA; 80, 100, 120, 140 kVp)을 변화시켜 심부선량과