Current Stimulator with Adaptive Supply Regulator for Artificial Retina Prosthesis

적응형 가변 전원 레귤레이터를 내장한 인공 망막용 전류 자극기

  • Ko, Hyoung-Ho (Department of Electronics, Chungnam National University)
  • 고형호 (충남대학교 전자공학과)
  • Received : 2011.03.24
  • Accepted : 2011.07.14
  • Published : 2011.07.31

Abstract

In this paper, a current stimulator circuit with adaptive supply regulator for retinal prosthesis is proposed. In current stimulation systems, the stimulating circuits with wide voltage swing range are needed due to the high impedance of the retina cell and microelectrodes. Thus, previous researches adopt the high voltage architecture to obtain the enough operating range. The high voltage architecture, however, could increase the power consumption and can damage the retina cells. The proposed circuit provides the adaptively regulated supply voltage by measuring the difference between desired stimulation current and the actual stimulation current. The proposed circuit can achieve the extended range of the allowable cell impedance, improved accuracy of the stimulation current, and higher biosafety.

Keywords

References

  1. Valerie C. Scanlon, Essentials of Anatomy and Physiology, 5th edition, F. A. Davis Company, Philadelphia, pp. 203-206, 2007.
  2. Ortmanns, M., Rocke, A., Gehrke, M., Tiedtke, and H.-J., "A 232-channel epiretinal stimulator ASIC," Solid-State Circuits, IEEE Journal of, vol. 42, no. 12, pp. 2946-2959, 2007. https://doi.org/10.1109/JSSC.2007.908693
  3. In Sook Kim, Jong Keun Song, Yu Lian Zhang, Tae Hyung Lee, Tae Hyung Cho, Yun Mi Song, Do Kyun Kim, Sung June Kim, and Soon Jung Hwang, "Biphasic electric current stimulates proliferation and induces VEGF production in osteoblasts," Biochimica et Biophysica Acta (BBA)-Molecular Cell Research, vol. 1763, no. 9, pp. 907-916, 2006.
  4. Wentai Liu, Kasin Vichienchom, Mark Clements, Stephen C. DeMarco, Chris Hughes, Elliot McGucken, Mark S. Humayun, Eugene de Juan, James D. Weiland, and Robert Greenberg, "A neuro-stimulus chip with telemetry unit for retinal prosthetic device," IEEE Journal of Solid-State Circuits, vol. 35, no. 10, pp. 1487-1497, 2000. https://doi.org/10.1109/4.871327
  5. Kuanfu Chen, Zhi Yang, Linh Hoang, James Weiland, Mark Humayun, and Wentai Liu, "An integrated 256-channel epiretinal prosthesis," IEEE Journal of Solid-State Circuits, vol. 45, no. 9, pp. 1946-1956, 2010. https://doi.org/10.1109/JSSC.2010.2055371
  6. Tokuda, T., Hiyama, K., Sawamura, S., Sasagawa, K., Terasawa, Y., Nishida, K., Kitaguchi, Y., Fujikado, T., Tano, Y., and Ohta, J., "CMOS-based multichip networked flexible retinal stimulator designed for image-based retinal prosthesis," Electron Devices, IEEE Transactions on, vol. 56, no. 11, pp. 2577-2585, 2009. https://doi.org/10.1109/TED.2009.2030552
  7. Theogarajan, L.S., "A low-power fully implantable 15-channel retinal stimulator chip," IEEE Journal of Solid-State Circuits, vol. 43, no. 10, pp. 2322-2337, 2008. https://doi.org/10.1109/JSSC.2008.2004331
  8. Bomin Kwon, Jinwoo Jung, Jiman Kim, Yongsu Park, and Hanjung Song, "Design of the low noise CMOS LDO regulator for a low power capacitive sensor interface," J. Kor. Sensors Soc., vol. 19, no. 1, pp. 25-30, 2010. https://doi.org/10.5369/JSST.2010.19.1.025