DOI QR코드

DOI QR Code

Development of Linux based Real-Time Spectrum Analyzer for Puretone Audiometer

순음청력검사기를 위한 리눅스 기반 실시간 스펙트럼 분석기 개발

  • Kang, Deok-Hun (Department of Biomedical Engineering, College of Natural Resource and Life Science, Pusan National University) ;
  • Shin, Bum-Joo (Department of Biomedical Engineering, College of Natural Resource and Life Science, Pusan National University) ;
  • Jeon, Gye-Rok (Department of Biomedical Engineering, School of Medicine, Pusan National University) ;
  • Wang, Soo-Geun (Department of Otolaryngology, School of Medicine, Pusan National University)
  • 강덕훈 (부산대학교 생명자원과학대학 바이오메디컬공학과) ;
  • 신범주 (부산대학교 생명자원과학대학 바이오메디컬공학과) ;
  • 전계록 (부산대학교 의학전문대학교 의공학교실) ;
  • 왕수건 (부산대학교 의학전문대학원 이비인후과학교실)
  • Received : 2011.04.29
  • Accepted : 2011.06.09
  • Published : 2011.06.30

Abstract

Because the puretone audiometer is an important medical instrument used to diagnose hearing loss, the IEC and ANSI has been published a specification with which audiometer should comply. This paper describes development of Linux based real-time spectrum analyzer which is dedicated to puretone audiometer. It can measure not only hearing level but also compliance of IEC standard for puretone audiometer such as frequency accuracy, harmonic distortion, pulsed tone, narrow band noise and linearity. We have verified our real-time spectrum analyzer through comparing to commercial product.

순음청력검사기는 청력의 손실 여부를 측정하기 위해 사용되는 중요한 의료기기이기 때문에 IEC와 ANSI에서는 순음청력검사기와 관련된 표준을 제시하고 있다. 본 논문은 순음청력검사기가 표준을 준수하는지 검사하기 위해 개발한 리눅스 기반 실시간 스펙트럼 분석기에 대해 기술한다. 본 논문의 실시간 스펙트럼 분석기는 청력 레벨 측정뿐 아니라 IEC에서 규정하고 있는 주파수 정확도, 청력 레벨 선형성, 고조파 왜곡, 단속음, 협대역 잡음에 대한 규정 준수 여부를 판단할 수 있는 기능을 제공한다. 개발한 실시간 스펙트럼 분석기를 상용 보정 시스템과 비교하여 검증한다.

Keywords

References

  1. K. S. Kim, "Guidelines on the Pure-tone Audiometry", KOSHA, October, 2006.
  2. ANSI "ANSI S1.4-1983 (R2006), American National Standard Specification for Sound Level Meters", 2006.
  3. Standards Secretariat Acoustical Society of America, "ANSI S3.6-2004, American National Standard Specification for audiometers", American National Standards Institute, May, 2004.
  4. TC 43, "ISO 389-1:l998, Acoustics - Reference zero for the calibration of audiometer equipment - Part 1 : Reference equivalent threshold sound pressure levels for pure tone and supra-aural earphones", International Organization for Standardization, 1998.
  5. IEC TC/SC 29, "IEC 60645-1:2001, Electroacoustics - Audiological Equipment - Part 1: Pure-Tone Audiometers", International Electrotechnical Commission, 2001.
  6. KS, "KS C IEC 60645-1:2005, Audiometers - Part 1 : Pure-tone audiometers", Korean Standards Association, December, 2005.
  7. K. S. Kim, et al., "Acoustic Calibration of Clinical Audiometers Used for Special Periodic Health Examination", Korean J Occup Environ Med, Vol. 16, No. 4, pp.381-390, December, 2004.
  8. Fons Adriaensen, "Audio Measurements using JAAA", LAC04 Presentation, Germany, 2004.
  9. Brian Phelps and Andreas Schwab, "BRP-PACU", http://sourceforge.net/projrcts/brp-pacu
  10. Nicholas J. Humfrey, "jackmeter", https://github.com/njh/jackmeter/
  11. Larson Davis, "Audiometer Calibration System", http://www.larsondavis.com/
  12. AUSSCO, "ACS100", http://audiometercal.com
  13. Paul Laumann et al., "Jack: The Jack Audio Connection Kit", http://jackaudio.org
  14. The Open Group, "IEEE Std 1003.1-2004, The core of the single UNIX Specification", http://www.unix.org/version3/ieee_std.html, April, 2004.
  15. ANSI, "ANSI S1.40-R2006, American National Standard Specifications and verification procedures for sound calibrators", 2006.
  16. EPL 29, " IEC 60942-2003, Electroacoustics - Sound calibrators", October, 2003.
  17. ANSI, "ANSI S3.1-1999 (R2008), Maximum Permissible Ambient Noise Levels for Audiometric Test Rooms", 2008.
  18. IEC, "IEC 60318-3:1998-08, Electroacoustics - Simulators of human head and ear - Part 3: Acoustic coupler for the calibration of supra-aural earphones used in audiometry", 1998.
  19. ANSI "ANSI S3.7-1995 (R2003), American National Standard Method for Coupler Calibration of Earphones", 1995.
  20. IEC "IEC 60651, The new International Standard for Sound Level Meter"
  21. IEC, "IEC 61094-4, Measurement microphones - Part 4: Specifications for working standard microphones", 1995.
  22. ANSI, "ANSI S1.40-2006, Verification Procedures for Sound Calibrators", 2006.
  23. IEC, "IEC 60942-2003, Electroacoustics - Sound calibrators", 2003.
  24. J. D. Kim, et al., "Development of Automatic Calibration System for PC-Based Pure Tone Audiometer", KAIS, Vol. 11, No. 7, pp.2586-2594, July, 2010. https://doi.org/10.5762/KAIS.2010.11.7.2586
  25. D. H. Kang, et al., "Development of SISI Test Software based on PC", KAIS, 제11권, 제4호, pp.1325-1332, April, 2010. https://doi.org/10.5762/KAIS.2010.11.4.1325
  26. B. D. Song, et al., "Development of PC-Based Client/Server Puretone Audiometer supporting Multiple Subject Concurrently", KAIS, Vol. 11, No. 6, pp.2225-2233, June, 2010. https://doi.org/10.5762/KAIS.2010.11.6.2225