Reliability Measurement of the Ultrasound for Safety of the Abdominal Organs on Acupuncturing Sangwan (CV13)

상완혈 자침시 복부 장기의 안전성 확보를 위한 초음파시스템의 신뢰도 측정

  • Kim, Ji-Hye (Dept. Biofunctional Medicine & Diagnostics, College of Korean Medicine, Sangji University) ;
  • Baek, Tae-Hyeun (Dept. of Internal Medicine, College of Korean Medicine, Sangji University)
  • 김지혜 (상지대학교 한의과대학 진단생기능의학교실) ;
  • 백태현 (상지대학교 부속한방병원 내과학교실)
  • Received : 2011.07.21
  • Accepted : 2011.08.12
  • Published : 2011.08.31

Abstract

Objectives: The objective of the current study is to determine whether an ultrasound device system is adequate for measuring distance from the abdominal surface of Sangwan (CV13) to peritoneum in front of anterior surface of liver. Methods: We recruited 3 healthy young male subjects and 2 sonographers. The each sonographer measured vertical shortest distance from the abdominal surface of Sangwan (CV13) to peritoneum in front of anterior surface of liver with a ultrasound device with three methods of 3.5 MHz convex probe and gel, 9 MHz linear probe and gel, and 9 MHz linear probe and solid gel pad, three times in random order. Because the total variation could be divided into repeatability, reproducibility and subject-to-subject variation in Gage R&R method, we compared the sources of variation associated with the measurement system with an analysis of variance model. Results & Conclusions: Number of distinct categories is calculated on the basis of standard deviation of subject-to-subject divided by standard deviation of total Gage R&R. If the number of categories is five or more, the measurement system may be acceptable for the analysis of the process. The number of distinct categories of the ultrasound device system for measuring distance from the abdominal surface of Sangwan (CV13) to peritoneum in front of anterior surface of liver were 9.20 (9 MHz linear probe and gel), 14.70 (9 MHz linear probe and solid gel pad). So we concluded that repeatabilities and reproducibilities of the ultrasound device system for measuring distance from the abdominal surface of Sangwan (CV13) to peritoneum in front of anterior surface of liver with the methods of 9 MHz linear probe and gel, and 9 MHz linear probe and solid gel pad were acceptable.

Keywords

References

  1. Lou XF, Yang XD, Jiang SH, Sun CY, Zhang RF. Study on angle and depth of needle insertion in acupuncture at Zusanli (ST36). Zhongguo Zhen Jiu. 2006; 26: 483-486.
  2. Li YD, Yang SD, Li JN, Li L. Study on needling depth of Dazhui (GV14), Jianzhongshu (SI15), Xuanshu (GV5) and Mingmen (GV4) with CT. Zhongguo Zhen Jiu. 2005; 25: 863-864.
  3. Yang SD, Li YD, Jiang GH, Hu NN, Cong SY. Study on needling depth of Fengfu (GV16) with CT. Zhongguo Zhen Jiu. 2008; 28: 47-48.
  4. Lin JG. Studies of needling depth in acupuncture treatment. Chin Med J (Engl). 1997; 110: 154-156.
  5. Chen HN, Lin JG, Yang AD, Chang SK. Safe depth of abdominal acupoints in pediatric patients. Complement Ther Med. 2008; 16: 331-335. https://doi.org/10.1016/j.ctim.2008.05.004
  6. Chen HN, Lin JG, Ying LC, Huang CT, Lin CH. The therapeutic depth of abdominal acupuncture points approaches the safe depth in overweight and in older children. J Altern Complement Med. 2009; 15: 1033-1037. https://doi.org/10.1089/acm.2008.0582
  7. 심찬섭. 복부초음파 진단학. 서울, 여문각, 2000, 9-10.
  8. 송한덕. 초음파진단의 이해. 서울, 군자출판사, 2000, 4.
  9. 전국한의과대학 침구․경혈학교실. 침구학 上. 서울, 집문당, 1994, 732-733.
  10. 전국한의과대학․한의학전문대학원 경락경혈학 교재편찬위원회. 대학경락경혈학각론 下. 원주, 의방출판사, 2009, 1179-81.
  11. 이승훈. Minitab 측정시스템 분석. 서울, 이레테크, 2006, 53-62.
  12. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of measurement. Lancet. 1986; 1: 307-310.