Recanalization of an Accidentally Crushed Coronary Stent by Intravascular Ultrasonography Catheter Entrapment

  • Kim, Jae-Yun (Division of Cardiology, Department of Internal Medicine, Soonchunhyang University College of Medicine) ;
  • Lee, Nae-Hee (Division of Cardiology, Department of Internal Medicine, Soonchunhyang University College of Medicine) ;
  • Cho, Yoon-Haeng (Division of Cardiology, Department of Internal Medicine, Soonchunhyang University College of Medicine) ;
  • Suh, Jon (Division of Cardiology, Department of Internal Medicine, Soonchunhyang University College of Medicine) ;
  • Seo, Hye-Sun (Division of Cardiology, Department of Internal Medicine, Soonchunhyang University College of Medicine) ;
  • Kim, Do-Hoi (Division of Cardiology, Department of Internal Medicine, Soonchunhyang University College of Medicine) ;
  • Jeon, Young-Woo (Division of Cardiology, Department of Internal Medicine, Soonchunhyang University College of Medicine)
  • Published : 2011.06.30

Abstract

Intravascular ultrasonography (IVUS) imaging is a user-friendly technique widely used during coronary interventions. An 80-year-old man was admitted with chest pain, and successful percutaneous coronary intervention was performed with stent implantation. One week later, the patient complained of further chest pain. Urgent coronary angiography showed total occlusion of the middle left anterior descending artery and the aspiration of thrombi was high. IVUS imaging showed inadequate stent strut apposition and distal dissection. We attempted another stent implantation but the IVUS catheter was stuck on the 0.014 inch wire. Therefore, we tried to pass the wire across the lateral side. After the wire was successfully passaged, the sprinter balloon was passed through the crushed stent to expand it. After 4 days later, the patient was discharged with no symptoms or electrocardiographic change.

Keywords

References

  1. Nicosia A, van der Giessen WJ, Airiian SG, von Birgelen C, de Feyter PJ, Serruys PW. Is intravascular ultrasound after coronary stenting a safe procedure?: three cases of stent damage attributable to ICUS in a tantalum coil stent. Cathet Cardiovasc Diagn 1997;40:265-70. https://doi.org/10.1002/(SICI)1097-0304(199703)40:3<265::AID-CCD9>3.0.CO;2-8
  2. Colombo A, Hall P, Nakamura S, et al. Intravascular stenting without anticoagulation accomplished with intravascular ultrasound guidance. Circulation 1995;91:1676-88. https://doi.org/10.1161/01.CIR.91.6.1676
  3. Bae JH, Rihal CS, Lerman A. Tissue characterization of coronary plaques using intravascular ultrasound/virtual histology. Korean Circ J 2006;36:553-8.
  4. Hausmann D, Erbel R, Alibelli-Chemarin MJ, et al. The safety of intra-coronary ultrasound: a multicenter survey of 2207 examinations. Circulation 1995;91:623-30. https://doi.org/10.1161/01.CIR.91.3.623
  5. Sasseen BM, Burke JA, Shah R, et al. Intravascular ultrasound catheter entrapment after coronary artery stenting. Catheter Cardiovasc Interv 2002;57:229-33. https://doi.org/10.1002/ccd.10244
  6. Batkoff BW, Linker DT. Safety of intracoronary ultrasound: data from a multicenter European registry. Cathet Cardiovasc Diagn 1996;38: 238-41. https://doi.org/10.1002/(SICI)1097-0304(199607)38:3<238::AID-CCD3>3.0.CO;2-9
  7. Alfonso F, Flores A, Escaned J, et al. Pressure wire kinking, entanglement, and entrapment during intravascular ultrasound studies: a potentially dangerous complication. Catheter Cardiovasc Interv 2000;50:221-5. https://doi.org/10.1002/(SICI)1522-726X(200006)50:2<221::AID-CCD18>3.0.CO;2-I

Cited by

  1. Successful surgical removal of an entrapped intravascular ultrasonography catheter in the left circumflex coronary artery vol.60, pp.12, 2011, https://doi.org/10.1007/s11748-012-0094-3
  2. Renal Stent Crushed During Open Aneurysmorrhaphy for Endoleak After Fenestrated EVAR vol.23, pp.3, 2016, https://doi.org/10.1177/1526602816642199
  3. A Case of Surgical Removal of an Intravascular Ultrasonography Catheter Entrapped in a Coronary Stent after Percutaneous Coronary Intervention vol.49, pp.6, 2011, https://doi.org/10.4326/jjcvs.49.362
  4. Successful Removal of an Entrapped Intravascular Ultrasound Catheter Using a 0.010-inch Guidewire and Compatible Balloon Catheter System vol.27, pp.1, 2011, https://doi.org/10.7793/jcad.27.20-00019