DOI QR코드

DOI QR Code

임플란트 식립시 발생한 열에 의한 골괴사: 증례보고

Thermally induced bone necrosis during implant surgery: 3 case reports

  • 조규홍 (인제대학교 의과대학 상계백병원 구강악안면외과학 교실) ;
  • 윤규호 (인제대학교 의과대학 상계백병원 구강악안면외과학 교실) ;
  • 박관수 (인제대학교 의과대학 상계백병원 구강악안면외과학 교실) ;
  • 배정호 (인제대학교 의과대학 상계백병원 구강악안면외과학 교실) ;
  • 유경하 (인제대학교 의과대학 상계백병원 구강악안면외과학 교실) ;
  • 한지훈 (인제대학교 의과대학 상계백병원 구강악안면외과학 교실) ;
  • 신재명 ;
  • 백지선 ;
  • 전인성 ;
  • 정정권 (인제대학교 의과대학 상계백병원 구강악안면외과학 교실)
  • Jo, Kyu-Hong (Department of Oral and Maxillofacial Surgery, Sanggye Paik Hospital, College of Medicine, Inje University) ;
  • Yoon, Kyh-Ho (Department of Oral and Maxillofacial Surgery, Sanggye Paik Hospital, College of Medicine, Inje University) ;
  • Park, Kwan-Soo (Department of Oral and Maxillofacial Surgery, Sanggye Paik Hospital, College of Medicine, Inje University) ;
  • Bae, Jung-Ho (Department of Oral and Maxillofacial Surgery, Sanggye Paik Hospital, College of Medicine, Inje University) ;
  • You, Kyung-Ha (Department of Oral and Maxillofacial Surgery, Sanggye Paik Hospital, College of Medicine, Inje University) ;
  • Han, Ji-Hoon (Department of Oral and Maxillofacial Surgery, Sanggye Paik Hospital, College of Medicine, Inje University) ;
  • Shin, Jae-Myung (Department of Oral and Maxillofacial Surgery, Ilsan Paik Hospital, College of Medicine, Inje University) ;
  • Baik, Jee-Seon (Department of Oral and Maxillofacial Surgery, Ilsan Paik Hospital, College of Medicine, Inje University) ;
  • Jeon, In-Seong (S & H Dental Clinic) ;
  • Cheong, Jeong-Kwon (Department of Oral and Maxillofacial Surgery, Sanggye Paik Hospital, College of Medicine, Inje University)
  • 투고 : 2011.05.30
  • 심사 : 2011.09.29
  • 발행 : 2011.10.31

초록

Thermally induced bone necrosis during implant surgery is a rare phenomenon and a potential contributing factor to implant failure. The frictional heat generated at the time of surgery causes a certain degree of necrosis of the surrounding differentiated and undifferentiated cells. The bone necrosis occurred in the mandible in all three cases, leading to a soft tissue lesion and pain. In each case, radiolucent areas appeared in the middle and apical portions of the implant 4 weeks after surgery. Thermally induced bone necrosis did not improve following systemic antibiotic medication, necessitating surgical treatment. The nonintegrated implants were removed, and meticulous debridement of dead bone and granulation tissue was performed. Then, new implants were implanted along with the placement of autogenous and xenogenic bone covered with a collagen membrane. No further complications occurred after re-operation. The radiolucencies around the new implants gradually resolved entirely, and the soft tissue lesions healed successfully. At 4-5 months after reoperation, implant loading was initiated and the implant-supported restorations have been functioning. The aim of this case report is to present the successful clinical treatment of three cases suspected to be caused by thermally induced bone necrosis after implant drilling.

키워드

참고문헌

  1. Branemark PI, Adell R, Breine U, Hansson BO, Lindstrom J, Ohlsson A. Intra-osseous anchorage of dental prostheses. I. Experimental studies. Scand J Plast Reconstr Surg 1969;3:81- 100. https://doi.org/10.3109/02844316909036699
  2. Higuchi KW, Folmer T, Kultje C. Implant survival rates in partially edentulous patients: A 3-year prospective multicenter study. J Oral Maxillofac Surg 1995;53:264-268. https://doi.org/10.1016/0278-2391(95)90222-8
  3. Mazzocchi A, Passi L, Moretti R. Retrospective analysis of 736 implants inserted without antibiotic therapy. J Oral Maxillofac Surg 2007;65:2321-2323 https://doi.org/10.1016/j.joms.2007.06.620
  4. Li W, Chow J, Hui E, Lee PKH, Cjow R. Resprospective study on immediate functional loading of edentulous maxillas and mandibles with 690 implants, up to 71 months of follow-up. J Oral Maxillofac Surg 2009;67:2653-2662 https://doi.org/10.1016/j.joms.2009.07.015
  5. Buser D, Schenk RK, Steinemann S, Fiorellini JP, Fox CH, Stich H. Influence of surface characteristics on bone integration of titanium implants. A histomorphometric study in miniature pigs. J Biomed Mater Res 1991;25:889-902. https://doi.org/10.1002/jbm.820250708
  6. Pye AD, Lockhart DEA, Dawson MP, Murray CA, Smith AJ. A review of dental implants and infection. J of Hospital Infection 2009;72:104-110 https://doi.org/10.1016/j.jhin.2009.02.010
  7. Bashutski JD, D`Silva NJ, Wang HL. Implant compression necrosis: Current understanding and case report. J Periodontol 2009;80:700-704. https://doi.org/10.1902/jop.2009.080581
  8. Quirynen M, De Soete M, Van Steenberghe D. Infectious risks for oral implants: A review of the literature. Clin Oral Implants Res 2002;13:1-19. https://doi.org/10.1034/j.1600-0501.2002.130101.x
  9. Branemark PI, Zarb GA, Albrektsson T. Tissue integrated prostheses: Osseointegration in clinical dentistry. Chicago, IL: Quintessence 1985:129-143.
  10. Eriksson R, Albrektsson T. Temperature thresholds for heat induced bone tissue injury: A vital-microscopic study in the rabbit. J Prosthet Dent 1983;50:101-107. https://doi.org/10.1016/0022-3913(83)90174-9
  11. Eriksson R, Albrektsson T. The effect of heat on bone regeneration: An experimental study in the rabbit using the bone growth chamber. J Oral Maxillofac Surg 1984;42:705-711. https://doi.org/10.1016/0278-2391(84)90417-8
  12. Sharawy M, Misch CE, Weller N, Tehemer S. Heat generation during implant drilling: The significance of motor speed. J Oral Maxillofac Surg 2002;60:1160-1169. https://doi.org/10.1053/joms.2002.34992
  13. Chacon GE, Bower DL, Larsen PR, McGlumphy EA, Beck FM. Heat production by 3 implant drill systems after repeated drilling and sterilization. J Oral Maxillofac Surg 2006; 64:265-269. https://doi.org/10.1016/j.joms.2005.10.011
  14. Eriksson R, Adell R. Temperatures during for the placement of implants using the osseointegration technique. J Oral Maxillofac Surg 1986;44:4-7. https://doi.org/10.1016/0278-2391(86)90006-6
  15. Cordioli G, Majzoub Z. Heat generation during implant site preparation: An in vitro study. Int J Oral Maxillofac Implants 1997;12:186-193.
  16. Ercoli C, Funkenbusch PD, Lee HJ, Moss ME, Graser GN. The influence of drill wear on cutting efficiency and heat production during osteotomy preparation for dental implants: A study of drill durability. Int J Oral Maxillofac Implants 2004;19:335-349.
  17. Queiroz TP, Souza FA, Okamoto R, Margonar R, Pereira-Filho VA, Garcia Junior IR et al. Evaluation of immediate bone-cell viability and of drill wear after implant osteotomies: Immunohistochemistry and scanning electron microscopy analysis. J Oral Maxillofac Surg 2008;66:1233-1240. https://doi.org/10.1016/j.joms.2007.12.037
  18. Yoshida K, Uoshima K, Oda K, Maeda T. Influence of heat stress to matrix on bone formation. Clin. Oral Impl. Res. 2009;20:782-790. https://doi.org/10.1111/j.1600-0501.2009.01654.x
  19. Cem Sener B, Dergin G, Gursoy B, Kelesoglu E, Slih I. Effects of irrigation temperature on heat control in vitro at different drilling depths. Clin. Oral Impl. Res. 2009;20:294-298. https://doi.org/10.1111/j.1600-0501.2008.01643.x
  20. Misir AF, Sumer M, Yenisey M, Ergioglu E. Effect of surgical drill guide on heat generated from implant drilling. J Oral Maxillofac Surg 2009;67:2663-2668. https://doi.org/10.1016/j.joms.2009.07.056
  21. Piattelli A, Piattelli M, Mangano C, Scarano A. A histologic evaluation of eight cases of failed dental implants: Is bone overheating the most probable cause? Biomaterials 1998;19:683-690. https://doi.org/10.1016/S0142-9612(97)00172-5
  22. Piattelli A, Scarano A, Balleri P, Favero GA. Clinical and histologic evaluation of an active" Implant Periapical Lesion": A case report. Int J Oral Maxiilofac Implants 1998;13:713-716.
  23. Penarrocha-Diage M, Boronat-Lopez A, Garcia-Mira B. Inflammatory implant periapical lesion: Etiology, diagnosis, and treatment - presentation of 7 cases. J Oral Maxillofac Surg 2009;67:168-173. https://doi.org/10.1016/j.joms.2007.12.022
  24. Sverzut AT, Vitti Stabile GA, Moraes M, Mazzonetto R, Fernandes Moreira RW. The influence of tobacco on early dental implant failure. J Oral Maxillofac Surg 2008;66:1004-1009. https://doi.org/10.1016/j.joms.2008.01.032
  25. Piattelli A, Scarano A, Piattelli M. Abscess formation around the apex of a maxillary root form implant: Cinical and microscopical aspects. A case report. J Periodontol 1995;66:899-903. https://doi.org/10.1902/jop.1995.66.10.899
  26. Alsaadi G, Quirynen M, Komarek A, van Steenberghe D. Impact of local and systemic factors on the incidence of oral implant failures, up to abutment connection. J Clin Periodontol. 2007;34: 610-617. https://doi.org/10.1111/j.1600-051X.2007.01077.x
  27. Pommer B, Frantal S, Willer J, Posch M, Watzek G, Tepper G.. Impact of dental implant length on early failure rates: a metaanalysis of observational studies. J Clin Periodontol. 2011; 38:856-863. https://doi.org/10.1111/j.1600-051X.2011.01750.x
  28. Park JC, Ha SR, Kim SM, Kim MJ, Lee JB, Lee JH. A randomized clinical 1-year trial comparing two types of non-submerged dental implants. Clin Oral Implants Res. 2010;21:228-236. https://doi.org/10.1111/j.1600-0501.2009.01828.x
  29. Matthews LS, Hirsch C. Temperature measured in human cortical bone when drilling. J Bone joint Surg[Am] 1972;54:297-308. https://doi.org/10.2106/00004623-197254020-00008
  30. Rhinelander F, Nelson CL, Steward RD, Steward CL. Experimental reaming of the proximal femur and acrylic cement implantation. Vascular and histologic effects. Clin Orthop 1979;141:74-89.
  31. Turkyilmaz I, Sennerby L, McGlumphy EA, Tozum TF. Biomechanical aspects of primary implant stability: A human cadaver study. Clin Implant Dent Relat Res 2009;11:113-119. https://doi.org/10.1111/j.1708-8208.2008.00097.x
  32. Tehemar SH. Factors affecting heat generation during implant site preparation: A review of biologic observations and future considerations. Int J Oral Maxillofac Implants 1999;14:127-136.
  33. Lundskog J. Heat and bone tissue. An experimental investigation of the thermal properties of bone and threshold levels for thermal injury. Scand J Plast Reconstr Surg 1972;9:1-80.
  34. Lavelle C, Wedgwood D. Effect of internal irrigation on frictional heat generated from bone drilling. J Oral Surg 1980;38:499- 503.
  35. Sutter F, Krekeler G, Schwammberger AE, Sutter FJ. Atraumatic surgical technique and implant bed preparation. Quintessence Int 1992;23:811-816.

피인용 문헌

  1. Temperature Values Variability in Piezoelectric Implant Site Preparation: Differences between Cortical and Corticocancellous Bovine Bone vol.2016, pp.None, 2011, https://doi.org/10.1155/2016/6473680
  2. Application of finite element analysis to evaluate optimal parameters for bone/tooth drilling to avoid thermal necrosis vol.8, pp.1, 2011, https://doi.org/10.1080/23311916.2021.1876582
  3. Thermal exposure of implant osteotomies and its impact on osseointegration—A preclinical in vivo study vol.32, pp.6, 2011, https://doi.org/10.1111/clr.13729