DOI QR코드

DOI QR Code

Image Quality Improvement after Implementation of a CT Accreditation Program

  • Kim, You-Sung (Department of Radiology, Seoul St. Marys Hospital) ;
  • Jung, Seung-Eun (Department of Radiology, Seoul St. Marys Hospital) ;
  • Choi, Byung-Gil (Department of Radiology, Seoul St. Marys Hospital) ;
  • Shin, Yu-Ri (Department of Radiology, Seoul St. Marys Hospital) ;
  • Hwang, Seong-Su (Department of Radiology, St. Vincent Hospital) ;
  • Ku, Young-Mi (Department of Radiology, Uijeongbu St. Mary's Hospital) ;
  • Lim, Yeon-Soo (Department of Radiology, Bucheon St. Mary's Hospital) ;
  • Lee, Jae-Mun (Department of Radiology, Yeouido St. Mary's Hospital, School of Medicine, The Catholic University of Korea)
  • 투고 : 2010.02.11
  • 심사 : 2010.05.10
  • 발행 : 2010.10.01

초록

Objective: The purpose of this study was to evaluate any improvement in the quality of abdominal CTs after the utilization of the nationally based accreditation program. Materials and Methods: Approval was obtained from the Institutional Review Board, and informed consent was waived. We retrospectively analyzed 1,011 outside abdominal CTs, from 2003 to 2007. We evaluated images using a fill-up sheet form of the national accreditation program, and subjectively by grading for the overall CT image quality. CT scans were divided into two categories according to time periods; before and after the implementation of the accreditation program. We compared CT scans between two periods according to parameters pertaining to the evaluation of images. We determined whether there was a correlation between the results of a subjective assessment of the image quality and the evaluation scores of the clinical image. Results: The following parameters were significantly different after the implementation of the accreditation program: identifying data, display parameters, scan length, spatial and contrast resolution, window width and level, optimal contrast enhancement, slice thickness, and total score. The remaining parameters were not significantly different between scans obtained from the two different periods: scan parameters, film quality, and artifacts. Conclusion: After performing the CT accreditation program, the quality of the outside abdominal CTs show marked improvement, especially for the parameters related to the scanning protocol.

키워드

참고문헌

  1. American Colleges of Radiology (ACR). CT accreditation program requirement, http://www.acr.org/accreditation/computed/ct_reqs.aspx
  2. Korean Institute for Accreditation of Medial Image (KIAMI). http://www.kiami.or.kr/
  3. Im TH, Na DG. Annual report No.1 2004-2006, Seoul: Korean Institute for Accreditation of Medical Image (KIAMI), 2007:1- 28, 75-105
  4. Department of education, KIAMI. The workshop for the examiners of quality assurance program (2006), Seoul: KIAMI, 2006:141-177
  5. Park HJ, Jung SE, Lee YJ, Cho WI, Do KH, Kim SH, et al. Review of failed CT phantom image evaluations in 2005 and 2006 by CT accreditation program of the Korean Institute for Accreditation of Medial Image. Korean J Radiol 2008;9:354- 363 https://doi.org/10.3348/kjr.2008.9.4.354
  6. Park HJ, Jung SE, Lee YJ, Cho WI, Do KH, Kim SH, et al. The relationship between subjective and objective parameters in CT phantom image evaluation. Korean J Radiol 2009;10:490-495 https://doi.org/10.3348/kjr.2009.10.5.490

피인용 문헌

  1. Quality Assurance in Ultrasound Screening for Hepatocellular Carcinoma Using a Standardized Phantom and Standard Clinical Images : A 3-Year National Investigation in Korea vol.33, pp.6, 2010, https://doi.org/10.7863/ultra.33.6.985
  2. Intra- and interobserver reliability of gray scale/dynamic range evaluation of ultrasonography using a standardized phantom vol.33, pp.2, 2010, https://doi.org/10.14366/usg.13021
  3. Evaluating the impact of accreditation and external peer review vol.28, pp.8, 2010, https://doi.org/10.1108/ijhcqa-05-2014-0055
  4. Impact of accreditation on performance of healthcare organizations : A review of global studies vol.12, pp.1, 2010, https://doi.org/10.1108/ijqss-10-2018-0085
  5. Effect of Digital Mammography for Breast Cancer Screening: A Comparative Study of More than 8 Million Korean Women vol.294, pp.2, 2020, https://doi.org/10.1148/radiol.2019190951