Abstract
The aim of this work was to establish the methodology for event positioning by measuring depth of interaction (DOI) information and to evaluate the system sensitivity and spatial resolution of the new detector for I-125 and Tc-99m imaging. For this purpose, a Monte Carlo simulation tool, DETECT2000 and GATE were used to model the energy deposition and light distribution in the detector and to validate this approach. Our proposed detector module consists of a monolithic CsI(Tl) crystal with dimensions of $50.0{\times}50.0{\times}3.0\;mm^3$. The results of simulation demonstrated that the resolution is less than 1.5 mm for both I-125 and Tc-99m. The main advantage of the proposed detector module is that by using 3 mm thick CsI(Tl) with maximum-likelihood position-estimation (MLPE) method, high resolution I-125 imaging and high sensitivity Tc-99m imaging are possible. In this paper, we proved that our new detector to be a reliable design as a detector for a multi-energy SPECT.
본 연구에서는 I-125 (35 keV) 와 Tc-99m (140 keV)에 대해 고해상도를 지닌 SPECT 영상을 동일한 검출기로 획득하는 방법을 제시하였고, 이를 몬테칼로 시뮬레이션 코드인 DETECT2000과 GATE를 이용하여 검증하였다. 제안된 검출기는 두께 3.0 mm의 CsI (Tl) 블록형 섬광체와 하마마츠사의 H8500C PSPMT로 이루어져 있다. 35 keV 감마선을 방출하는 I-125 핵종을 영상화할 때는 두꺼운 섬광체를 사용 할 경우, 일반적인 앵거방법으로는 빛 퍼짐이 많아지기 때문에 내인성 공간분해능이 저하되지만, 최대우도 함수와 색인테이블을 사용하여 감마선 반응 위치를 추적하면 내인성 공간분해능을 향상시킬 수 있다. DETECT2000 시뮬레이션 결과 1.0 mm 이내의 내인성 공간분해능을 획득하였다. 140 keV를 방출하는 Tc-99m를 영상화할 경우에는 I-125 전용인 1.0 mm 두께의 섬광체를 사용하였을 경우보다 3.0 mm 두께의 섬광체를 사용하였을 때 2.3배 이상 향상된 민감도를 보였다. 본 연구에서 제안한 검출기의 장점은 저에너지의 I-125 선원에 대해 상대적으로 두꺼운 섬광체를 사용하더라도 최대우도함수를 사용하기 때문에 분해능의 감소가 없다는 점과 Tc-99m 선원에 대해 민감도의 저하가 적다는 점이다. 본 연구에서 제안한 검출기를 사용하면 고에너지와 저에너지를 가진 핵종을 모두 영상화할 수 있다.