Abstract

Ankle-foot orthosis with an artificial pneumatic muscle which is intended for the assistance of plantarfelxion torque was developed. In this study, power pattern of the device in the various pneumatics and the effectiveness of the system were investigated. The pneumatic power was provided by ankle-foot orthosis controlled by user‘s physiological signal, that is, muscular stiffness in soleus muscle. This pneumatic power can assist plantarflexion torque of ankle joint. The subjects performed maximal voluntary isokinetic plantarflexion motion on a biodexdynamometer in different pneumatics, and they completed three conditions: 1) without wearing the orthosis, 2) wearing the orthosis with artificial muscles turned off, 3) wearing the orthosis activated under muscular stiffness control. Through these experiments, we confirmed the effectiveness of the orthosis and muscular stiffness control using the analyzing isokinetic plantarflexion torque. The experimental results showed that isokinetic torques of plantarflexion motion of the ankle joints gradually increased in incremental pneumatic. The effectiveness of the orthosis was -7.26% and the effectiveness of the muscular stiffness control was 17.83% in normalized isokinetic plantarflexion torque. Subjects generated the less isokinetic torques of the ankle joints in wearing the orthosis with artificial muscles turned off, but isokinetic torques were appropriately reinforced in condition of wearing the orthosis activated under muscular stiffness control(17.83%) compared to wearing the orthosis(-7.26%). Therefore, we respect that developed powered orthosis is applied in the elderly that has weak muscular power as the rehabilitation equipment.

Keywords

References

  1. Perry, J., "Kinesiology of lower extremity bracing," Clinical Orthopaedics and Related Research, Vol. 102, pp. 18-31, 1974. https://doi.org/10.1097/00003086-197407000-00004
  2. Bobath, B., "Adult hemiplegia; Evaluation & treatment (3rd ed.)," Butterworth-Heinemann, pp. 80-85, 1990.
  3. Carr, J. H. and Shepherd, R. B., "Physiotherapy in disorders of the brain," Petrocelli Books, pp. 95-110, 1980.
  4. Ahn, S. C., Hwang, S. J., Kang, S. J. and Kim, Y. H., "Development of a portable gait phase detection system for patients with gait disorders," J. Biomed. Eng. Res., Vol. 26, No. 3, pp. 145-150, 2005.
  5. Hwang, S. J., Kim, J. Y., Hwang, S. H., Park, S. W., Yi, J. B. and Kim, Y. H., "Development of the Active Ankle Foot Orthosis to Induce the Normal Gait for the Paralysis patients," Journal of the Ergonomics Society of Korea, Vol. 26, No. 2, pp. 131-136, 2007. https://doi.org/10.5143/JESK.2007.26.2.131
  6. Ferris, D. P., Czerniecki, J. M. and Hannaford, B., "An ankle-foot orthosis powered by artificial pneumatic muscles," Journal of Applied Biomechanics, Vol. 21, No. 2, pp. 189-197, 2005.
  7. Gordon, K. E., Sawichi, G. S. and Ferris, D. P., "Mechanical performance of artificial pneumatic muscles to power an ankle-foot orthosis," Journal of Biomechanics, Vol. 39, No. 10, pp. 1832-1841, 2006. https://doi.org/10.1016/j.jbiomech.2005.05.018
  8. Yamamoto, S., Ebina, M., Kubo, S., Hayashi, T., Akita, Y. and Hayakawa, Y., "Development of an ankle-foot orthosis with dorsiflexion assist. Part 2: structure and evaluation," J. Prosthet. Orthot., Vol. 11, No. 2, pp. 24-28, 1999.
  9. Yokoyama, O., Sashika, H., Hakiwara, A., Yamamoto, S. and Yasui, T., "Kinematic Effects on Gait of a Newly Designed Ankle-Foot Orthosis with Oil Damper Resistance: A case Series of 2 patients With Hemiplegia," Archives of Physical Medicine and Rehabilitation, Vol. 86, No. 1, pp. 162-166, 2005. https://doi.org/10.1016/j.apmr.2003.11.026
  10. Sakaguchi, M. and Furusho, J., "Force Display System Using Particle-Type Electrorheological Fluids," Proceedings of the 1998 IEEE International Conference on Robotics and Automation, pp. 2586-2590, 1998.
  11. Sasaki, D., Noritsugu, T. and Takaiwa, M., "Development of Pneumatic Power Assist Splint "ASSIST" Operated by Human Intention," Journal of Robotics and Mechatronics, Vol. 17, No. 5, pp. 568-574, 2005. https://doi.org/10.20965/jrm.2005.p0568
  12. Norris, J. A., Granata, K. P., Mitros, M. R., Byrne, E. M. and Marsh, A. P., "Effect of augmented plantarflexion power on preferred walking speed and economy in young and older adults," Gait & Posture, Vol. 25, No. 4, pp. 620-627, 2007. https://doi.org/10.1016/j.gaitpost.2006.07.002
  13. Kao, P. C., Lewis, C. L. and Ferris, D. P., "Invariant ankle moment patterns when walking with and without a robotic ankle exoskeleton," Journal of Biomechanics, Vol. 43, No. 2, pp. 203-209, 2010. https://doi.org/10.1016/j.jbiomech.2009.09.030
  14. Perry, J., "Gait Analysis-Normal and pathological Function," SLACK, Inc., pp. 54-62, 1992.
  15. Choi, Y. C., Rhee, K. M. and Choi, H. S., "The Stress Distribution Property on the Customized Ankle Foot Orthoses During the Gait Period," J. of the Korea Society for Precision Engineering, Vol. 25, No. 3, pp. 165-175, 2008.
  16. Perry, J. and Burnfield, J., "Gait Analysis: Normal and Pathological Function," SLACK, Inc., pp. 53-91, 1992.
  17. Kim, K., Kang, S. R., Piao, Y. J., Jeong, G. Y. and Kwon, T. K., "Analysis of the Assist Characteristics for Torque of the Ankle Plantarflexion in Elderly Adults Wearing the Ankle-Foot Orthosis," The Journal of Korea Robotics Society, Vol. 5, No. 1, pp. 48-54, 2010.
  18. Gordon, K. E. and Ferris, D. P., "Learning to walk with a robotic ankle exoskeleton," Journal of Biomechanics, Vol. 40, No. 12, pp. 2636-2644, 2007. https://doi.org/10.1016/j.jbiomech.2006.12.006