Abstract
Quality management of an x-ray unit drastically differs according to the type of establishment of medical institutions. Many primary medical institutions have it, but they do not pay much attention to quality management. In the study, Gyeongbuk area has been divided into four zones from January 4, 2010 to September 3, 2010, and four places were designated by city. Among medical institutions located at a total of 16 sites, the target was 8 places with X-ray emission equipment 10 years or more in use as well as 8 places with X-ray emission equipment less than 10 years in use. The 5 essential items of quality control were tested. In the test that checked for equipment it was found that sites with X-ray emission equipment 10 years or more in use didn't have ground connection (6.25%) while all of them passed the current leak test. In the exposure dose reproducibility test 4 sites with X-ray emission equipment 10 years or more in use (25%) and 1 site with X-ray emission equipment less than 10 years in use didn't pass the test. In the KVp accuracy test 5 sites with X-ray emission equipment 10 years or more in use (31.25%) and 2 sites with X-ray emission equipment less than 10 years in use (12.5%) didn't pass the test. In the tube current and tube current amount test 3 sites with X-ray emission equipment 10 years or more in use (18.75%) and 1 site with less than 10 years in use (6.25%) didn't pass the test. According to the findings of the present research, quality control at medical institutions with X-ray equipment 10 years or more in use was poorer than medical institutions with X-ray equipment less than 10 years in use. In this regard, administrative and technical measures need to be taken as soon as possible. In addition, owners of medical clinics or unit managers need to raise awareness, and it is necessary to revise a regular test cycle every year or every two years if they have old equipment or if the equipment is not used on a frequent basis. And it is also important to provide regular educational programs for quality management.
X선 발생장치의 정도관리는 의료기관의 설립형태에 따라 가장 큰 차이가 있으며, 장치의 보유율은 1차 의료기관에서 가장 많이 보유한 반면 정도관리에 대한 인식과 관리는 소홀하다 할 수 있다. 본 연구에서는 2010년 1월 4일부터 2010년 9월 3일까지 경북 지역을 4개의 권역으로 나누고, 각 시별로 4곳을 임으로 선정하였다. 전체 16곳에 있는 의원급 의료기관 중 10년 이상 된 X선 발생장치를 보유한 8곳과 10년 이하의 X선 발생장치를 보유한 8곳를 그 대상으로 하였으며, 정도관리의 필수 항목 5가지를 시험하였다. 접지 설비 확인시험에서는 10년 이상 된 X선 장치를 보유한 한 곳이 별도의 접지선을 구비하지 않았으며(6.25%), 외장 누설전류시험에서는 모두 적합 하였다. 조사선량 재현성 시험에서는 10년 이상 된 X선 발생장치를 보유한 4곳(25%)과 10년 이하의 X선 발생장치를 보유한 1(6.25%)곳이 부적합 하였다. 관전압의 정확도 시험에서는 10년 이상 된 X선 발생장치를 보유한 5곳(31.25%)과 10년 이하의 X선 장치를 보유한 2곳(12.5%)이 부적합 하였다. 관전류와 관전류량 시험에서는 10년 이상 된 X선 장치를 보유한 3곳(18.75%)과 10년 이하의 X선 장치를 보유한 1곳(6.25%)이 부적합 하였다. 본 연구 결과 10년 이하의 X선 발생장치를 보유한 의원급 의료기관 보다 10년 이상 된 의원급 의료기관의 X선 발생장치의 정도 관리 실태는 심각한 수준 이였다. 검사 결과를 바탕으로 행정적, 기술적인 조치들이 신속하게 이루어져야 할 것이다. 또한 의원급의 사업주나 장치 담당자들의 의식 전환이 필요 하며 그러기 위해서는 오래된 장비이거나 자주 사용하지 않는 장비일수록 정기검사주기를 1~2년 단위로 개정하여야 하고, 정기적인 정도 관리 교육 프로그램을 실시하여야 할 것으로 사료된다.