Study for the Indirect Measuring Method of Operational Force in Surgical Robot Instrument

복강경 수술용 로봇 인스트루먼트의 간접적 작동력 측정법에 관한 연구

  • Received : 2010.06.10
  • Accepted : 2010.07.20
  • Published : 2010.09.01

Abstract

This paper proposes the method indirectly measuring the operating force of the end-effect tip of surgical robot instrument which conducts the surgical operation in the body on behalf of the surgeon's hand. Due to the size and safety obligation to the surgical robot instrument, it is difficult to measure the operation force of its tip like grasping force. However the instrument is driven by cable-pulley torque transmission mechanism and when some force is occurred at the tip, then the reaction force appears on the cable as additional tension. Based on this phenomenon, this paper proposes a method to estimate the operating force from measuring reaction force against the driving motor by using a loadcell. And it induces mathematical equation to calculate the force from loadcell by approaching the modulus of elasticity to high order polynomial. And this paper proves the validity of proposed mechanism by experimental test.

Keywords

References

  1. Laparoscopy Modular Instruments of XION Medical GmbH, http:// www.xion-medical.com.
  2. P. Giulianotti, A. Coratti, M. Angelini, F. Sbrana, S. Cecconi, T. Balestracci, and G. Caravaglios, “Robotics in general surgerypersonal experience in a large community hospital,” Archives of Surgery, no. 238, pp. 777-784, 2003.
  3. C. Preusche, T. Ortmaier, and G. Hirzinger, “Teleoperation concepts in minimal invasive surgery,” Control of Engineering Practice, vol. 10, no. 11, pp. 1245-1250, November 2002. https://doi.org/10.1016/S0967-0661(02)00084-9
  4. EndoWrist Instruments Catalogue in Intuitive Surgical, Inc, http://www.intuitivesurgical.com
  5. J. Bodner, H. Wykypiel, G. Wetscher, and T. Schmid, “First experiences with the da VinciTM operating robot in thoracic surgery,” European Journal of Cardio-Thoracic Surgery, no. 25, pp. 844-851, 2004.
  6. L. M. Aksman, C. R. Carignan, and D. L. Akin, “Force estimation based compliance control of harmonically driven manipulators,” Proc. of presented at the ICRA 2007, Roma, Italy, pp. 4208-4213, April 2007.
  7. T. K. Lee, M. C. Lee, and C. Y. Kim, “Modeling of torque transmission characteristics in surgical robot,” ICROS-SICE International Joint Conference 2009, pp. 2617-2620, August 18-21, 2009
  8. G. Fu, F. Moses, and D. A. Khazem, “Strength of parallel wire cables for suspension bridges,” in 8th ASCE Specialty Conference on Probabilistic Mechanics and Structural Reliability, pp. 222-227, July 22-25, 2000
  9. Medical Cable Assemblies Brochure, Carl Stahl Sava Industries, Inc, http://www.savacable.com.
  10. Cable Design Catalog, Carl Stahl Sava Industries, Inc Web Site, Available: www.savacable.com.

Cited by

  1. Design and evaluation of a slave manipulator with roll-pitch-roll wrist and automatic tool loading mechanism in telerobotic surgery vol.8, pp.4, 2012, https://doi.org/10.1002/rcs.1458
  2. Dynamic modeling of coupled tendon-driven system for surgical robot instrument vol.15, pp.10, 2014, https://doi.org/10.1007/s12541-014-0566-8