The Korean Journal of Nuclear Medicine Technology (핵의학기술)
- Volume 14 Issue 2
- /
- Pages.208-211
- /
- 2010
- /
- 1229-9901(pISSN)
- /
- 2982-8406(eISSN)
Evaluation of Coated Tube CA19-9 IRMA kit
Coated tube를 사용한 CA19-9 측정용 IRMA 시약의 평가
- Lee, Hyun-Ju (Department of Nuclear Medicine, Seoul Metropolitan Government Seoul National University Boramae Medical Center) ;
- Jang, Hyun-Young (Department of Nuclear Medicine, Seoul Metropolitan Government Seoul National University Boramae Medical Center) ;
- Shin, Sun-Young (Department of Nuclear Medicine, Seoul Metropolitan Government Seoul National University Boramae Medical Center) ;
- Kim, Hee-Sun (Department of Nuclear Medicine, Seoul Metropolitan Government Seoul National University Boramae Medical Center) ;
- Kim, Tae-Hoon (Department of Nuclear Medicine, Seoul National University Hospital) ;
- Lee, Ho-Young (Department of Nuclear Medicine, Seoul Metropolitan Government Seoul National University Boramae Medical Center)
- 이현주 (서울특별시보라매병원 핵의학과) ;
- 장현영 (서울특별시보라매병원 핵의학과) ;
- 신선영 (서울특별시보라매병원 핵의학과) ;
- 김희선 (서울특별시보라매병원 핵의학과) ;
- 김태훈 (서울대학교병원 핵의학과) ;
- 이호영 (서울특별시보라매병원 핵의학과)
- Received : 2010.09.02
- Accepted : 2010.09.27
- Published : 2010.10.30
Abstract
Purpose:
현재 본원은 TFB사의 CA19-9 항체가 bead에 코팅되어있는 CA19-9 IRMA 시약을 사용하고있다. 이번에 TFB사에서 새로 개발된 CA19-9 항체가 polystyrene test tube에 코팅된 CA 19-9 IRMA 시약이 기존 TFB사의 bead를 사용하는 시약을 대체하여 사용할 수 있는지 시약의 성능을 평가하였다. 본원과 서울대학교병원 환자 56명의 검체를 대상으로 하여 TFB사의 기존 bead 방법을 사용한 시약과 새로 개발된 coated tube를 사용한 시약을 비교하여 상관계수와 회귀식을 구하고 coated tube를 사용한 시약의 정밀도, 회수율, 직선성, 민감도, Hook effect 를 확인하였다. 저, 중, 고역가 검체의 검사내 정밀도는 4.1%, 4.0%, 4.2%이고 검사간 정밀도는 7.6%, 4.3%, 7.8%이며 회수율은 모두