Breast Specific Gamma Imaging 장비의 성능평가

Performance Evaluation of Breast Specific Gamma Imaging

  • 차은선 (삼성서울병원 핵의학과) ;
  • 곽인석 (삼성서울병원 핵의학과) ;
  • 노익상 (삼성서울병원 핵의학과) ;
  • 연준호 (삼성서울병원 핵의학과) ;
  • 김기 (삼성서울병원 핵의학과) ;
  • 최춘기 (삼성서울병원 핵의학과) ;
  • 석재동 (삼성서울병원 핵의학과)
  • Cha, Eun-Sun (Department of Nuclear Medicine, Samsung Medical Center) ;
  • Kwak, In-Suk (Department of Nuclear Medicine, Samsung Medical Center) ;
  • Noh, Ik-Sang (Department of Nuclear Medicine, Samsung Medical Center) ;
  • Yeon, Joon-Ho (Department of Nuclear Medicine, Samsung Medical Center) ;
  • Kim, Ki (Department of Nuclear Medicine, Samsung Medical Center) ;
  • Choi, Choon-Ki (Department of Nuclear Medicine, Samsung Medical Center) ;
  • Seok, Jae-Dong (Department of Nuclear Medicine, Samsung Medical Center)
  • 투고 : 2010.01.06
  • 심사 : 2010.01.26
  • 발행 : 2010.06.05

초록

Breast cancer는 조기 병변의 발견이 매우 중요하나 크기가 작은 암세포를 구별해 내는 데 한계가 있어 예민도 96.4% 이상의 성능을 얻을 수 있는 BSGI가 도입되어 임상적인 중요성이 강조되고 있다. 따라서 예민한 감도를 보이는 BSGI 장비를 이용하여 계수율, 해상력 등의 성능 평가를 하고자 한다. Dilon사의 BSGI 장비와 $^{99m}Tc$ 점 선원 1.85 MBq (0.05mCi)부터 1.85~37 (0.05~1 mCi) MBq 간격으로 111 MBq (4mCi)까지 이용하였다. 성능 평가 방법은 BSGI 콜리메이터 유효시야(useful field of view, UFOV) 5개의 지점을 선택하여 60초씩 영상을 얻어 평균 count 변화로 불응시간을 측정하였다. 또한 Breast에 임상적 섭취 분포를 기준으로 콜리메이터 유효시야에서 점 선원을 1.85, 3.7, 5.55, 7.4 MBq (0.05, 0.1, 0.15, 0.2 mCi)까지 증가시켜 점 선원과의 거리를 10, 20, 30, 40, 50 mm 간격으로 FWHM을 얻었다. $^{99m}Tc$ 74 MBq (2mCi) 이하에서 count는 증가하고 그 이상에서는 더 이상 증가 하지 않는다. 점 선원의 거리에 의한 FWHM에서 1.85 MBq (0.05 mCi)는 4.05, 4.73, 5.77, 6.90, 8.00, 9.32 mm, 3.7 MBq (0.1 mCi)는 4.30, 4.80, 5.90, 7.00, 8.10, 9.07 mm, 5.55 MBq (0.15 mCi)는 4.90, 5.60, 6.20, 7.20, 8.20, 9.10 mm, 7.4 MBq (0.2 mCi)는 5.30, 6.10, 6.60, 7.00, 7.90, 8.70 mm로 거리에 따라 FWHM 약 0.5-1 mm의 변화를 얻었다. BSGI 장비의 검출기 자체의 특성인 불응시간으로 인하여 74 MBq (2 mCi) 이상 사용으로는 왜곡된 영상을 얻을 수 있다는 것을 알았다. 또한 FWHM은 거리, 일정한 분포 변화, 병변위치에 따라 검출기를 다양한 각도에서 최대한 밀착하게 하면 더 유용하게 검사할 수 있다.

Purpose: Early diagnosis of breast is of the utmost importance to improve prognosis. We have a limitation for mammography and sonography detecting small cancer. Clinical importance of Breast Specific Gamma Imaging (BSGI) has improved for that reason. So We studied performance evaluation test of count rate and resolution with high sensitivity to the low dose of BSGI. Materials and Methods: BSGI of Dilon 6800, point source of $^{99m}Tc$ from 1.85~148 MBq (0.05~4 mCi) at the intervals of 1.85~37 MBq (0.05~1 mCi) was used for the test. Performance evaluation method was performed for measuring deadtime for choosing at the 5 different point in the useful field of view (UFOV), acquired image for 60 seconds. Compared with reference of clinical uptake distribution of breast, activity increased according to the distance change 10, 20, 30, 40, 50 mm in the useful field of view. Results: Counting curve increased according to the activity from 1.85 MBq (0.05 mCi) to the 74 MBq (2 mCi), and it change flat shape over 74 MBq (2 mCi). The variation of the full width of half maximum (FWHM) to the distance is 4.05, 4.73, 5.77, 6.90, 8.00, 9.32 mm in 1.85 MBq (0.05 mCi), 4.30, 4.80, 5.90, 7.00, 8.10, 9.07 mm in 3.7 MBq (0.1 mCi), 4.90, 5.60, 6.20, 7.20, 8.20, 9.10 mm in 5.55 MBq (0.15 mCi), 5.30, 6.10, 6.60, 7.00, 7.90, 8.70 mm in 7.40 MBq (0.2 mCi). Conclusion: Distortions of image would be acquired because of the deadtime in BSGI. We found out the fact that specification of $^{99m}Tc$ reaction under 74 MBq (2 mCi) for BSGI. Second, FWHM distribution change from varied distance from the detector, clearly distinguished the location of the lesion.