The Optimization of Reconstruction Method Reducing Partial Volume Effect in PET/CT 3D Image Acquisition
PET/CT 3차원 영상 획득에서 부분용적효과 감소를 위한 재구성법의 최적화
- Hong, Gun-Chul (Department of Nuclear Medicine, Samsung Medical Center) ;
- Park, Sun-Myung (Department of Nuclear Medicine, Samsung Medical Center) ;
- Kwak, In-Suk (Department of Nuclear Medicine, Samsung Medical Center) ;
- Lee, Hyuk (Department of Nuclear Medicine, Samsung Medical Center) ;
- Choi, Choon-Ki (Department of Nuclear Medicine, Samsung Medical Center) ;
- Seok, Jae-Dong (Department of Nuclear Medicine, Samsung Medical Center)
- 홍건철 (삼성서울병원 핵의학과) ;
- 박선명 (삼성서울병원 핵의학과) ;
- 곽인석 (삼성서울병원 핵의학과) ;
- 이혁 (삼성서울병원 핵의학과) ;
- 최춘기 (삼성서울병원 핵의학과) ;
- 석재동 (삼성서울병원 핵의학과)
- Received : 2010.02.02
- Accepted : 2010.02.25
- Published : 2010.06.05
Abstract
Purpose: Partial volume effect (PVE) is the phenomenon to lower the accuracy of image due to low estimate, which is to occur from PET/CT 3D image acquisition. The more resolution is declined and the lesion is small, the more it causes a big error. So that it can influence the test result. Studied the optimum image reconstruction method by using variation of parameter, which can influence the PVE. Materials and Methods: It acquires the image in each size spheres which is injected
PVE는 PET/CT 3D 영상 획득에서 발생되는 것으로 평가값이 저평가되어 영상의 정확도를 떨어뜨리는 현상이다. 특히 이는 병소의 크기가 작고 분해능이 저하될수록 더 큰 오차를 초래하여 검사 결과에 영향을 줄 수 있다. 본 연구는 PVE에 영향을 줄 수 있는 매개변수의 변화를 이용하여 최적의 영상 재구성법을 알아보고자 한다. GE Discovery STE16 장비에서 NEMA 2001 IEC phantom을 이용하여 각기 다른 크기의 구체(직경 37, 28, 22, 17, 13, 10 mm)에
Keywords
- Partial volume effect (PVE);
- Iteration;
- Subset number;
- Standard uptake value (SUV);
- Signal to noise ratio (SNR);
- % Difference