DOI QR코드

DOI QR Code

Calculation Method of Entrance Skin Dose in X-ray Beam Quality Factor

선질계수에 의한 피부입사선량 계산법

  • 김성철 (가천의과학대학교 방사선학과) ;
  • 김종일 (전북대학교 과학학과) ;
  • 안성민 (가천의과학대학교 방사선학과)
  • Published : 2010.02.28

Abstract

This interest in radiation exposure makes increasing doctor's awareness and knowledge of radiation dose in patients during X-ray test important in reducing patient's uneasiness. However, very few facilities are equipped with measurement instruments. Therefore, an intensive study to find out patient dose using computational method has been initiated. This study used special features of the bit system and NDD-M and directly measured the output dose of diagnostic X-ray instruments used in Korea to create tables. Two different methods were found to be adequate when applied to cases when X-ray outputs were both known and unknown, and comparative experiments with real measurement doses were carried out. Presented methods were found to provide more accurate results compared to the bit system and NDD-M. Therefore, patient dose during clinical trials were found to be more easily acceptable to medical personnel in the radiation field in terms of radiation exposure and reduction of medical X ray dose.

방사선피폭에 대한 관심이 높아지면서 X선 검사시에 환자에게 조사되는 피폭선량을 정확히 알고 있다는 것은 환자의 불안을 해소하고 또 방사선사나 의사가 피폭선량 경감의식을 향상시키는 데 중요하지만, 임상에서 측정기를 보유하고 있는 시설은 극소수에 불과하다. 본 연구에서는 bit system 및 NDD-M법의 특징을 살려서 우리나라에 사용되고 있는 진단용 X선장치의 출력선량을 직접 측정하여 도표화 하고, X선 출력선량을 아는 경우 또는 모르는 경우 모두에서 적절히 적용할 수 있게 두 가지 방법을 제시하여 실측선량과 비교 실험을 하였다. 그 결과 bit system 및 NDD-M법보다 정확도가 우수한 결과를 나타내어 임상에서 환자가 받는 선량을 더욱 쉽게 알 수 있게 됨으로 방사선관련 종사자들의 의료피폭에 대한 관심이 더욱 높아지고 의료선량감소에 한층 더 노력하는 계기가 될 것으로 사료된다.

Keywords

References

  1. E. Stephen Amis, Priscilla F. Butler, KimberlyE. Applegate, Steven B. Brinbaum, Libby F.Brateman, James M. Hevezi, Fred A. Mettler,Richard L. Morin, Michael J. Pentecost,Geoffrey G. Smith, Keith J. Strauss and RobertK. Zeman, "American college of radiology whitepaper on radiation dose in medicine," Journal ofthe american college of radiology, Vol.4, No.5,pp.272-284, 2007. https://doi.org/10.1016/j.jacr.2007.03.002
  2. 임봉식, "한국에서 방사선관련 종사자들의 개인 피폭선량 실태에 관한 연구", 방사선기술과학, 제29권, 제3호, pp.185-195, 2006.
  3. ICRP, Based on International Commission onRadiological Protection, ICRP Publication 92,2004.
  4. UNSCEAR, Sources and Effects of IonizingRadiation. (Report to the General Assembly)(New York : United Nations), 2000.
  5. 조평곤, CT진단 검사를 위한 최적선량 준위에 관한 연구, 고려대학교대학원, 2007.
  6. IAEA, International basic safety standards forprotection against Ionizing radiation and forthe safety of radiation sources. IAEA safetyseries No. 115, 1996.
  7. ICRP supporting Guidance 2, Diagnosticreference levels in medical imaging : Reviewand additional advice. ICRP committee 3,2002.
  8. NRPB, National protocol for patient dosemeasurements in diagnostic radiology, IPSM,1992.
  9. 식품의약품안전청, 환자선량측정 가이드라인. 방사선안전관리시리즈 No.14, 2007.
  10. 김우란, 이춘식, 이재기, “진단X선에 의한 성인의 진단행위별 유효선량평가”, 방사선방어학회지, 제27권, 제3호, pp.155-164, 2002.
  11. 김정민, 석진용, “촬영조건에 의한 환자 피폭선량의 자동계산 프로그램”, 대한방사선기술학회지, 제25권, 제2호, pp.35-38, 2002.
  12. 김정민, 허준, 임태랑, 석전유치, 앵정달야, “흉부 X선 사진농도로 부터 표면선량을 산출하는 방법”, 대한방사선기술학회지, 제22권, 제1호, pp.13-20, 1999.
  13. E. I. DuPont de Nemours & Co., Inc, The Bitsystem of Technique Management, 8th. ed.,Wilmington. Delaware.
  14. Japan society of radiological technologyIbaragi branch office, Non-Dosimeter-Dosimetry(NDD) Method. Japan socity ofradiological technology Ibaragi branch office.Mito. 1996.
  15. 진단용방사선발생장치의 안전관리에 관한 규칙, 보건복지가족부령 제1호, 2008.
  16. 한국표준협회, KS A 4021, 의료용 X선장치 통칙, 1997.
  17. 한국표준협회, KS A ISO 4037-1, 광자에너지에 따른 응답결정과 선량계 및 선량율계 교정용 엑스, 감마 기준방사선-제1부:방사선특성과 발생방법, 2003.
  18. Host aichinger, Joachim dierker, Sigrid$joite-barfu{\beta}$, and Manfred sabel, RadiationExposure and Image Quality in X-rayDiagnostic Radiology, Springer, 2004.
  19. 김유현, 최종학, 김성수, 이찬엽, 조평곤, 이영배, 김철민, “진단방사선영역에서 방사선장치의 이용실태 및 환자피폭선량에 관한 조사연구”, 의학물리, 제16권, 제1호, pp.10-15, 2005.
  20. AAPM report No. 31, Standardized Methodsfor Measuring Diagnostic X-ray Exposure,1990.
  21. 한준구, 박재형, 강위생, 한만청, 김주완, “진단방사선과 촬영실에서의 방사선 피폭”, 대한방사선의학회지, 제26권, 제2호, pp.422-427, 1990.
  22. 김정민, 김성철, 최종학, 고신관, 이인자, 김선칠, 신귀순, 양한준, 방사선영상정보학, 신광출판사, 2008.

Cited by

  1. Comparative Study of Radiation Exposure using Entrance Skin Dose Calculation Technique in Diagnostic X-Ray Radiography vol.11, pp.12, 2011, https://doi.org/10.5392/JKCA.2011.11.12.357