Abstract
This paper is about a projection method to check existence of radius groove penetration of screw after distal radius fracture operation using the T-type plate. Angle of Radius groove was analyzed by fifty one CT images that contains patients' wrist and twenty cases of radius specimens. After making radius phantom by plaster, we set the screw so that it penetrated 2.4 mm depth of radius groove. Then, we projected the phantom by X-ray in change of the elevation and supination angle of distal radius by 5 degree interval on 0~30 degree. The average value of groove angle in the wrist CT images was 14.4 degree and the radius specimens was 16.3 degree. Screws penetrating radius groove of the phantom have different lengths according to elevation angle and supination angle. Consequently, in order to confirm existence and nonexistence of radius groove penetration of the screw in tangential projection after distal radius fracture operation using the T-type plate, we recommend 5 degree of elevation angle and 20 degree of supination angle.
T형 금속판을 이용한 요골 원위부 골절 수술 후 나사못의 요골구 관통 유무를 확인하기 위한 요골구 촬영 방법을 알아보고자 하였다. Wrist CT 환자 51명의 3D 영상과 건조된 요골 표본 20본에서 요골구 각도를 분석하였고, 석고를 이용하여 요골 모형을 제작 한 다음 나사못이 요골구를 2.4 mm 관통 하도록 하고 요골구 원위부 거상 각도 및 외회전 각도를 0~30도까지 5도씩 변화시켜 X-선 영상으로 촬영하였다. 그 결과 요골구 각도는 Wrist CT 영상에서 평균 14.4도, 건조된 요골 표본에서 평균 16.3도였다. 요골 모형에서 요골구를 관통한 나사못의 길이는 거상 각도 및 외회전 각도에 따라 묘출 정도가 다름을 알 수 있었다. 결론적으로 T형 금속판을 이용한 요골 원위부 골절 수술 후 나사못의 요골구 관통을 알기 위한 접선방향촬영에서 요골 원위부 거상각은 $5^{\circ}$, 외회전 각은 $20^{\circ}$를 추천한다.