A Comparision of the Radiation dose by Distance and the Direction according to a Tube Position of the C-arm Unit
C-arm의 Tube 위치에 따른 거리 및 방향별 피폭선량 비교
- Kim, Jin-Su (Department of Radiology, Asan Medical Center) ;
- Woo, Bong-Cheol (Department of Radiology, Asan Medical Center) ;
- Kim, Sung-Jin (Department of Radiology, Asan Medical Center) ;
- Lee, Kwan-Sup (Department of Radiology, Asan Medical Center) ;
- Ha, Dong-Yoon (Department of Radiology, Asan Medical Center)
- 김진수 (서울아산병원 영상의학팀) ;
- 우봉철 (서울아산병원 영상의학팀) ;
- 김성진 (서울아산병원 영상의학팀) ;
- 이관섭 (서울아산병원 영상의학팀) ;
- 하동윤 (서울아산병원 영상의학팀)
- Published : 2009.05.30
Abstract
In operation room, the use of the C-arm unit is increasing. So, the radiation dose of the person who work in operation room was even more increased than before. Thus, this study is shown the measurement of expose dose and the way for decrease of the radiation dose by using the C-arm unit. The experiment was performed with the C-arm unit and used a phantom which is similar to tissue of the human body and fluoro-glass dosimeter for dose measurement. The expose dose were measured by the tube position(over tube, under tube) of the C-arm unit, distance(50, 100