Usefulness Assessment of Automatic Analysis Program for Flangeless Esser PET Phantom Images

Flangeless Esser PET Phantom 영상 자동 분석 프로그램의 유용성 평가

  • NamGung, Chang-Kyeong (Department of Nuclear Medcine, Asan Medical Center) ;
  • Nam, Ki-Pyo (Department of Nuclear Medcine, Asan Medical Center) ;
  • Kim, Kyeong-Sik (Department of Nuclear Medcine, Asan Medical Center) ;
  • Kim, Jeong-Seon (Department of Nuclear Medcine, Asan Medical Center) ;
  • Lim, Ki-Cheon (Department of Nuclear Medcine, Asan Medical Center) ;
  • Shin, Sang-Ki (Department of Nuclear Medcine, Asan Medical Center) ;
  • Cho, Shee-Man (Department of Nuclear Medcine, Asan Medical Center) ;
  • Dong, Kyung-Rae (Department of Radiological Technology1, Gwangju Health College University)
  • Received : 2009.03.11
  • Accepted : 2009.04.02
  • Published : 2009.05.11

Abstract

Purpose: ACR (American College of Radiology) offers variable parameters to PET/CT quality control by using ACR Phantom. ACR Phantom was made to evaluate parameters which are uniformity, attenuation, scatter, contrast and resolution. Manual analysis method wasn't good for the use of QC because values of parameter were changed as it may user and it takes long time to analysis. Ki-Chun Lim, a nuclear scientist in AMC, developed program that automatically analysis values of parameter by using ACR Phantom to overcome above problems. In this study, we evaluated automatic analysis program's usability, through the comparing SUV of each method, reproducibility of SUV when repeated analysis and the time required. Materials and Methods: Using Flangeless Esser PET Phantom, the ideal ratio of 4 : 1 hot cylinder and BKG but it actually showed a ratio of 3.89 to 1 hot cylinder and BKG. SIEMENS Biograph True Point 40 was used in this study. We obtained images using ACR phantom at Fusion WB PET Scan condition (2 min/bed) and 120 kV, 100 mAs CT condition. Using True X method, 3 iterations, 14 subsets, Gaussian filter, FWHM 4 mm and Zoom Factor 1.0, $168{\times}168$ image size. We obtained Max. & Min. SUV and SUV Mean values at Cylinder (8, 12, 16, 25 mm, Air, Bone, Water, BKG) by automatic program and obtained SUV by manual method. After that, we compared manual and automatic method. we estimate the time required from opened the image data to final work sheet was completed. Results: Automatic program always showed same result and same the time required. At 8, 12, 16 and 25 m cylinder, manual method showed 6.69, 3.46, 2.59, 1.24 CV values. The larger cylinder size became, the smaller CV became. In manual method, bone, air, water's CV were over 9.9 except BKG (2.32). Obtained CV of Mean SUV showed BKG was low (0.85) and bone was high (7.52). The time required was 45 second, 882 second respectably. Conclusions: As a result of difference automatic method and manual method, automatic method showed always same result, manual method showed that the smaller hot cylinders became, the lager CV became. Hot cylinders mean region size, the smaller hot cylinder size becomes we had some trouble in doing ROI poison setting. And it means increase in variation of SUV. The Study showed the time required of automatic method was shorten then manual method.

Keywords