A Study on Treatment Target Position Verification by using Electronic Portal Imaging Device & Fractionated Stereotatic Radiotherapy

EPID와 FSRT를 이용한 치료표적위치 검증에 관한 연구

  • 이동훈 (동명대학교 의용공학과) ;
  • 권장우 (동명대학교 컴퓨터공학과) ;
  • 박승우 (원자력의학원 방사선의학연구소) ;
  • 김윤종 (원자력의학원 방사선의학연구소) ;
  • 이동한 (원자력의학원 방사선의학연구소) ;
  • 지영훈 (원자력의학원 방사선의학연구소)
  • Published : 2009.05.25

Abstract

It is very important to verify generated setup errors in cancer therapy by using a high energy radiation and to perform the precise radiation therapy. Specially, the verification of treatment position is very crucial in special therapies like fractionated stereotatic radiotherapy (FSRT). The FSRT uses normally high-dose, small field size for treating small intracranial lesions. To estimate the developed FSRT system, the isocenter accuracy of gantry, couch and collimator were performed and a total of inaccuracy was less than ${\pm}1mm$. Precise beam targeting is crucial when using high-dose, small field size FSRT for treating small intracranial lesions. The EPID image of the 3mm lead ball mounted on the isocenter with a 25mm collimator cone was acquired and detected to the extent of one pixel (0.76mm) after comparing the difference between the center of a 25mm collimator cone and a 3 mm ball after processing the EPID image. In this paper, the radiation treatment efficiency can be improved by performing precise radiation therapy with a developed video based EPID and FSRT at near real time

고 에너지 방사선을 이용하여 종양을 치료하는 과정 중 발생하는 오차를 확인하여 보다 정교한 치료를 시행하는 것은 매우 중요하다. 특히, 두경부 종양을 치료하기 위해 작은 조사면에 방사선을 집중 조사하는 정위 방사선 분할치료와 같은 특수 치료에서는 치료위치의 확인을 더욱 요구하고 있다. 작은 조사면에 고선량의 방사선을 표적체적에 집중 조사하는 정위 방사선 분할 치료법에서 치료 중심점의 정확도 검증은 매우 중요하므로 표적에 3mm 납 볼을 놓고 정위 방사선 분할 치료용 25mm 콘을 장착한 후 얻은 전자 포탈 영상으로부터 콘 원의 중심점과 볼 원의 중심점을 영상처리를 통해 서로 비교해 본 결과 1픽셀(0.76mm)의 정확도까지 얻을 수 있었다. 정위 방사선 분할 치료에 전자 포탈 영상 장치를 적용하여 치료위치오차 여부를 검증하였다. 이를 위해 첫 번째 정위 방사선 분할 치료 영상과 임의로 2mm 이동시킨 후 얻은 영상의 두개골 외곽선을 검출한 후 서로 비교해 본 결과 발생된 2mm 오차를 검증할 수 있었다. 본 논문에서는 개발된 비디오 기반 전자포탈영상장치와 정위 방사선 분할치료장치를 이용하여 실시간적으로 포탈영상을 획득할 수 있었으며 치료위치 검증을 통해 보다 정교한 방사선 치료를 할 수 있으리라 본다.

Keywords

References

  1. M. Goitein, J. Busse, 'Immobilization errors: Some theoretical considerations,' Radiology, Vol. 117, pp. 407-412, 1975 https://doi.org/10.1148/117.2.407
  2. A. Dutreix, 'When and how can we improve precision in radiotherapy?,' Radiother. Oncol. Vol. 2, pp. 275-292, 1984 https://doi.org/10.1016/S0167-8140(84)80070-5
  3. K. Mah, J. VanDyk, T. Keane, et al., 'Acute radiation-induced pulmonary damage: A clinical study on the response to fractionated radiotherapy,' Int. J. Radiat. Oncol. Biol. Phys., Vol. 13, pp. 179-188, 1987 https://doi.org/10.1016/0360-3016(87)90125-8
  4. K. Theodorou, C. Kappas, C. Tsokas., 'A new non-invasive and relocatable immobilization frame for fractionated stereotactic radiotherapy', Radiotherapy and Oncology, Vol. 47, pp. 313-317, 1998 https://doi.org/10.1016/S0167-8140(98)00015-2
  5. R. W. Laing, A. P. Warrington, F. Hines, et al., 'Fractionated stereotactic external beam radiotherapy in the management of brain metastasis,' Eur. J. Cancer, Vol. 29, pp. 1387-1391, 1993 https://doi.org/10.1016/0959-8049(93)90007-3
  6. R. J. Maciunas, R. L. Galloway, J. W. Latimer., 'The application accuracy of stereotactic frames,' Neurosurgery, Vol. 35(4), pp. 682-695, 1994 https://doi.org/10.1227/00006123-199410000-00015
  7. M. C. Schell, V. Smith, D. A. Larson, et al., 'Evaluation of radiosurgery techniques with cumulative dose volume histograms in linac-based stereotactic external beam irradiation,' Int. J. Radiation Oncology Biol. Phys., Vol. 20, pp. 1325-1330, 1990 https://doi.org/10.1016/0360-3016(91)90245-Y
  8. E. Shaw, R. Kline, M. Gillin, et al., 'radiosurgery quality assurance guidelines,' Int. J. Radiation Oncology Biol. Phys., Vol. 27, pp. 1231-1239, 1993 https://doi.org/10.1016/0360-3016(93)90548-A
  9. A. P. Warrington 'Quality assurance in fractionated stereotactic radiotherapy,' radiother. oncol. Vol. 30, pp. 239-246, 1994 https://doi.org/10.1016/0167-8140(94)90464-2
  10. C. F. Serago, 'Radiosurgery target point alignment errors detected with portal film verification,' Int. J. Radiation Oncology Biol. Phys., Vol. 24, pp. 777-780, 1992 https://doi.org/10.1016/0360-3016(92)90728-Z
  11. D. Lei, S. Almon, T. Samuel 'Verification of radiosurgery target point alignment with an elcetronic portal imaging device(EPID),' Med. Phy., Vol. 24(2), pp. 263-267, 1997 https://doi.org/10.1118/1.598070
  12. M. V. Herk, H. Meertens, 'A matrix ionisation chamber imaging device for on-line patient setup verification during radiotherapy,' Radiother. Oncol., Vol. 11, pp. 369-378, 1988 https://doi.org/10.1016/0167-8140(88)90208-3
  13. L. E. Antonuk, C. W. Kim, J. Boudry, et al. 'Development of hydrogenated amorphous silicon sensors for diagnostic x-ray imaging' IEEE Trans. Nuc. Sci., Vol. 38, pp. 636-640, 1991 https://doi.org/10.1109/23.289367
  14. A. C. Savard, E. B. Podgorsak, 'An electronic portal imaging device as a physics tool,' Medical Dosimetry, Vol. 22(2), pp. 101-105, 1997 https://doi.org/10.1016/S0958-3947(97)00002-2
  15. E. C. McCullough, K. P. McCollough, 'Improving agreement between radiation-delineated field edges on simulation and portal films : The edge tolerance test tool,' Med. Phys., Vol. 20, pp. 375-376, 1993 https://doi.org/10.1118/1.597129
  16. K. Luchka 'Assessing radiation and light field congruence with a video based electronic portal imaging device,' Med. Phy., Vol. 23(7), pp. 1245-1252, 1996 https://doi.org/10.1118/1.597867