Abstract
In this study, we measured the dose distribution of scattered ray in X-ray radiography room using an ion chamber and examined the dependency of scattered ray content on the scattered ray source and exposure condition. To study the factors of scattered ray occurrence in the acryl phantom, we measured the change in the scatted ray content according to the X-ray tube voltage (40~140 kV) and the field size ($10{\times}10\;cm^2$, $20{\times}20\;cm^2$, $35{\times}35\;cm^2$). For the $35{\times}35\;cm^2$ field size, the side-scattering rate ranged from 3.1% to 14.5%. The scattered ray contributions of the phantom, collimator, X-ray tube and wall were also measured. The scattered ray contribution of the phantom was higher than 95.4% for the entire tube voltage, and those of the collimator, X-ray tube and wall were 2.6%, 1.3% and 0.7%, respectively.
본 연구에서는 전리함을 이용하여 엑스선 촬영실 내에서 공간산란선량을 측정하고 산란선의 발생원과 조사 조건에 대한 산란율 의존성을 조사하였다. 아크릴 팬톰의 산란선 발생 인자를 조사하기 위해 관전압(40~140 kV)과 조사면($10{\times}10\;cm^2$, $20{\times}20\;cm^2$, $35{\times}35\;cm^2$)에 대한 산란율 변화를 측정하였다. 조사면이 $35{\times}35\;cm^2$일 때 측방산란율은 3.11~4.5%로 측정되었다. 촬영실 내에서 팬톰, 콜리메이터, 엑스선관, 벽면에서 발생하는 산란선량의 영향을 측정하였는데 전체 관전압에 대해 팬톰의 산란선 발생률은 95.4% 이상이었으며 콜리메이터, 엑스선관, 벽면의 산란선 발생률은 각각 2.6%, 1.3%, 0.7% 이하로 나타났다. 관전압 100 kV, 40 mAs 조사 조건에서 촬영실 내 팬톰으로부터 1 m 거리에서 측정한 공간선량은 최대 2 mR 정도로 나타나 조사조건의 최적화 등의 산란선 경감조치가 필요할 것으로 생각된다.