A Comparison of Dose in Changed Technique Factor Using X-ray Imaging System

X-선 장치의 기술적 인자의 변화에 따른 선량 비교 평가

  • Han, Dong-Kyoon (Department of Radilogical Science, Eulji University) ;
  • Ko, Shin-Gwan (Department of Radilogical Science, Eulji University) ;
  • Seon, Jong-Ryul (Department of Radiological Technology, Daejeon Health Science College) ;
  • Yoon, Seok-Hwan (Department of Radiological Technology, Dongnam Health Science College) ;
  • Jung, Jae-Eun (Department of Digital Management and Information, Nambu University)
  • 한동균 (을지대학교 보건과학대학 방사선학과) ;
  • 고신관 (을지대학교 보건과학대학 방사선학과) ;
  • 선종률 (대전보건대학 방사선과) ;
  • 윤석환 (동남보건대학 방사선과) ;
  • 정재은 (남부대학교 디지털경영정보학과)
  • Published : 2009.11.30

Abstract

With the recent development of diagnosis using radiation and increasing demand of the medical treatment, we need to minimize radiation exposure dose. So, This is the method which reduce patient dose by measuring surface dose of radiographic change factor and by comparing theoretical and actual dose, when we take an X-ray which is generally used. By changing the factor of kV, mAs, FSD, whose range is 60 to 120 kV, 20 to 100 mAs, 80 to 180 cm, we compared theoretical surface dose with actual surface dose calculated by the simple calculation program, Bit system, and NDD-M method As a result, when kV and mAs were higher, theoretical surface dose and actual surface dose were more increased. but the higher FSD was, the more decreased surface dose was. According to this, the error were measured about 0.1 to 0.2 mGy in low dose part and about 0.7 to 1.5 mGy in high dose part. Therefore, this shows that theoretical surface dose calculation method is more correct in low dose part than in high dose part. In conclusion, we will have to make constant efforts which can reduce patient and radiographer's exposure dose, studying methods which can predict patient's radiation exposure dose more exactly.

Keywords