DOI QR코드

DOI QR Code

Analysis of Properties and Phantom Design Based on Plastic Hardener and Softener for Ultrasonic Imaging

초음파 영상용 플라스틱 기반의 팬텀제작 및 특성 분석

  • Lee, G.J. (Department of Biomedical Engineering, Yonsei University) ;
  • Park, D.H. (Department of Biomedical Engineering, Yonsei University) ;
  • Shin, T.M. (Department of Biomedical Engineering, Yonsei University) ;
  • Seo, J.B. (Department of Biomedical Engineering, Yonsei University)
  • 이균정 (연세대학교 보건과학대학 의공학과) ;
  • 박동희 (연세대학교 보건과학대학 의공학과) ;
  • 신태민 (연세대학교 보건과학대학 의공학과) ;
  • 서종범 (연세대학교 보건과학대학 의공학과)
  • Published : 2008.08.30

Abstract

Plastic hardener and softener based ultrasound phantoms were made in various constitutions and their acoustic properties were measured. Speed of sound is approximately $1.4\;mm/{\mu}sec$ in all the phantoms, which is about 7% less than that of in soft tissue. Attenuation coefficient is strongly dependent on the ratio between hardener and softener. In order to achieve the tissue level attenuation (0.5 dB/cm/MHz), 60% of hardener or less is required. The synthesized phantoms can be preserved for more than 6 months without structural degradation.

Keywords

References

  1. E.L. Madsen, J.A. Zagzebski, R.A. Banjavie, and R.E. Jutila, "Tissue mimicking materials for ultrasound phantoms", Medical Physics, Vol. 5, No.5. Pp 391-394. 1978 https://doi.org/10.1118/1.594483
  2. M.M. Burlew, E.L. Madsen, J.A. Zagzebski, R.A. Banjavie, S.W. Sum, "A new ultrasound tissue-equivalent Material", Radiation Physics, Vol. 134, pp517-520, 1980
  3. T.L. Szabo, "Diagnostic Ultrasound Imaging", Elsevier, 2004
  4. K. Takegami , Y . Kaneko, T . Watanabe, T . Maruyama, Y . Matsumoto, H . Nagawa, "Polyacrylamide gel containing egg white as new model for irradiation experiments using focused ultrasound", Ultrasound in Medicine & Biology, Volume 30 , Issue 10, pp 1419 - 1422 https://doi.org/10.1016/j.ultrasmedbio.2004.07.016
  5. "Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment", AlUM, 2004
  6. RC. Preston, DR Bacon, and RA Smith, "Calibration of Medical Ultrasound Equipment-Procedures and Accuracy Assessement", IEEE transactions on UFFC, Vol. 35. No.2. Pp 110-121,1988 https://doi.org/10.1109/58.4161
  7. RF. Wagner, S.W. Smith, H.M. Sandrik, H. Lopez, "Statistics of Speckle in Ultrasound B-Scans", IEEE transactions on sonics and ultrasonics, Vol. 30, No.3, pp 156-163. 1983 https://doi.org/10.1109/T-SU.1983.31404
  8. E.L. Madsen, F. Dong, G.R. Frank, B.S. Garra, K.A. Wear, T. Wilson, J.A. Zagzebski, H.L. Miller, K.K. Shung, S.H. Wang, E.J. Feleppa, T. Liu, W.D. O'Brien Jr., K.A. Topp, N.T. Sanghiv, A.V. Zaitsev, T.J. Hall, J.B. Fowlkes, O.D. Kripfgans, J.G. Miller, "Interlaboratory Comparison of ultrasonic backscatter, attenuation, and speed measurements". J Ultrasound Med, Vol. 18,pp.615-631, 1999 https://doi.org/10.7863/jum.1999.18.9.615