Surface Dose Measurement of Electron Beam within the Magnetic Field Variation

자기장 내에서 전자선의 표면선량 변화 측정

  • 제재용 (인제대학교 의과대학 부산백병원 방사선종양학과) ;
  • 노경석 (마산대학 방사선과) ;
  • 신운재 (동의과학대학 방사선과) ;
  • 박철우 (동의과학대학 방사선과)
  • Received : 2008.05.28
  • Accepted : 2008.08.25
  • Published : 2008.09.30

Abstract

Purpose: This paper describes a electron field presence of magnetic field, intensity and shape surface dose variation to clinical application possibility. Materials and Methods: The using 6 MeV electron and $10{\times}10\;cm^2$ field size, 9 hole to shielding block make the by measure the film, when the magnetic field position inside and outside of the X-Omat film and parallel plate ionization chamber using the surface dose measured. Results: Present of 4 cm to the side at angle about 3 degree from beam center, use of ring type magnetic is 0.9% increase the surface dose, lens block located in the magnetic field the surface dose 1.58% increase, half magnetic field's position on the side of them at the field center of the 3.6% increase of the surface dose. Conclusion: Surface dose variation is with magnetic field about the mean electron beam of progress direction change, orbit region patient's is inconvenient without surface dose increase percentage case goodness will be used as a useful way.

목적: 전자선 조사야에서 자기장의 유무, 세기와 형태에 따른 표면 선량의 변화를 알아보고 임상에서의 적용 가능성에 대하여 알아보고자 한다. 대상 및 방법: 6 MeV 전자선과 $10{\times}10\;cm^2$의 조사야를 이용하여 9개의 hole이 있는 차폐블록을 제작하여 필름 농도를 측정하고 전자선 조사야 내에 자장이 위치 할 때와 조사야의 외부에 위치 할 때의 선량변화를 필름과 표면선량 측정기를 이용하여 측정하였다. 결 과: 전자선의 입사각은 조사야의 중심에서 측면으로 4 cm 지점에서 3도 정도의 입사각을 나타내었으며, ring type의 자기장을 사용할 경우 0.9%의 표면선량 증가가 나타나고, lens block에 자기장을 위치시키면 1.58%의 표면선량이 증가하고, half 조사야의 중심 측면에 자기장을 위치시키면 조사야의 중심에 3.6%의 표면 선량 증가를 보였다. 결 론: 전자선은 선속의 중심에서 측면으로 일정 각도를 이루고 조사되어지며 자기장이 위치한 조사야내 표면선량의 변화는 자기장으로 인하여 전자선의 진행 방향이 변화되었음을 의미하고 안구와 같은 부위에 환자의 불편함이 없이 표면선량을 증가 시켜야 할 경우 유용한 방법으로 사용되어질 것이다.

Keywords