The Usefulness Assessment of Attenuation Correction and Location Information in SPECT/CT

SPECT/CT에서 감쇠 보정 및 위치 정보의 유용성 평가

  • Choi, Jong-Sook (Department of Nuclear Medicine, Asan Medical Center) ;
  • Jung, Woo-Young (Department of Nuclear Medicine, Asan Medical Center) ;
  • Shin, Sang-Ki (Department of Nuclear Medicine, Asan Medical Center) ;
  • Cho, Shee-Man (Department of Nuclear Medicine, Asan Medical Center)
  • 최종숙 (서울아산병원 핵의학과) ;
  • 정우영 (서울아산병원 핵의학과) ;
  • 신상기 (서울아산병원 핵의학과) ;
  • 조시만 (서울아산병원 핵의학과)
  • Received : 2008.09.30
  • Accepted : 2008.10.16
  • Published : 2008.10.25

Abstract

Purpose: We make a qualitative analysis of whether Fusion SPECT/CT can find lesion's anatomical sites better than existing SPECT or not, and we want to show the usefulness of SPECT/CT through finding out effects of CT attenuation correction on SPECT images. Materials and Method: 1. The evaluation of fusion images: This study comprised patients who was tested $^{131}I$-MIBG, Bone, $^{111}In$-Octreotide, Meckel's diverticulum, Parathyroid MIBI with Precedence 16 or Symbia T2 from 2008 Jan to Aug. We compared SPECT/CT image with non fusion image and make a qualitative analysis. 2. The evaluation of attenuation correction: We classified 38 patients who was tested 201Tl myocardial exam with Symbia T2 into 5 sections by using Cedars Sinai' QPS program - Ant, Inf, Lat, Septum, Apex. And we showed each section's perfusion states by percentage. We compared the each section's perfusion-states differences between CT AC and Non AC by average${\pm}$standard deviation. Results: 1. The evaluation of fusion images : In high energy $^{131}I$ cases, it was hard to grasp exact anatomical lesions due to difference between regions and surrounding lesions' uptake level. After combining with CT, we could grabs anatomical lesion more exactly. And in meckel's diverticulum case or to find lesions around bowels or organs with $^{111}In$ cases, it demonstrates its superiority. Bone SPECT/CT images help to distinguish between disk spaces certainly and give correct results. 2. The evaluation of attenuation correction: There is no significant difference statistically in Ant and Lat (p>0.05), but there is a meaningful difference in Inferior, Apex and Septum (p<0.05). AC perfusion at inferior wall in the 5 sections of myocardium: The perfusion difference between Non AC perfusion image ($68.58{\pm}7.55$) and CT corrected perfusion image ($76.84{\pm}6.52$) was the largest by $8.26{\pm}4.95$ (p<0.01, t=10.29). Conclusion: Nuclear medicine physicians can identify not only molecular image which shows functional activity of lesions but also anatomical location information of lesions with more accuracy using the combination of SPECT and CT systems. Of course this combination helps nuclear medicine physician find out the abnormal parts. Moreover combined data sets help separate between normal group and abnormal group in complicated body part. So clinicians can carry out diagnosis and treatment planning at the same time with a single test image. In addition, when we examine a myocardium in thorax where attenuation can occur easily, we can trust perfusion more in a certain region in SPECT test because CT provides the capability for accurate attenuation correction. In these reasons, we think we can prove the justice after treatment fusion image.

목적 : 융합 SPECT/CT가 기존 SPECT에 비해 병소의 해부학적 위치를 정확히 판단할 수 있는지 정성 평가하고, CT 감쇠 보정이 SPECT 영상에 미치는 효과를 알아보아 SPECT/CT의 유용성을 제시하고자 한다. 실험재료 및 방법 : 1. 융합 영상의 평가 : 2008년 1월(月)부터 8월(月)까지 Precedence 16 혹은 Symbia T2에서 $^{131}I$-MIBG, Bone, $^{111}In$-Octreotide, Meckel 게실, Parathyroid MIBI 등을 SPECT/CT 시행한 환자를 대상으로 하였고, SPECT/CT영상을 융합한 것과 하지 않은 것을 비교하여 정성 평가하였다. 2. 감쇠보정의 평가 : Symbia T2로 2008년 6월에서 8월까지 $^{201}Tl$ 심근 검사를 한 환자 38명을 대상으로 Cedars-Sinai의 QPS를 이용하여 산출하였다. Ant, Inf, Lat, Septum, Apex로 5개부분으로 분류하고, 각 부분에 대한 관류의 상태를 백분율로 산출했다. CT AC와 Non AC를 평균${\pm}$표준편차로 각 부분에 대한 관류 상태를 비교하고 차이를 분석하였다. 결과 : 1. 융합 영상의 평가 : 에너지가 높은 $^{131}I$ SPECT의 경우 병소와 주위 조직 간의 섭취율 차이로 인해 (주위 조직이 saturation 됨) 병소의 위치 파악이 어려웠으나 CT로 융합한 결과 해부학적 위치를 정확히 평가할 수 있었다. 또한 멕켈게실이나 $^{111}In$과 같이 장이나 장기쪽에 질환을 찾는 경우에는 그 우수성이 더욱 뛰어 났다. Bone SPECT/CT는 척추간의 구별을 확실히 할 수 있어 임상의가 정확한 결과를 제시하는데 도움을 준다. 2. 감쇠 보정의 평가 : 감쇠 보정 전후의 관류 백분율의 차이가 Ant, Lat에서는 통계적으로 유의한 차이가 없었으나(p>0.05), Inferior, Apex, Septum에서는 유의한 차이가 있었다(p<0.05). 차이를 보이는 값 중 Inferior Wall에서 CT AC perfusion : $76.84{\pm}6.52%$, Non AC perfusion : $68.58{\pm}7.55%$로 CT 보정에 의한 차이가 $8.26{\pm}4.95%$로 가장 크게 측정되었다(t=10.29, p<0.01). 결론 : SPECT에 CT가 부착되면서 병변의 기능적 활성도를 나타내는 분자학적 영상은 물론 병변의 해부학적 위치 정보를 보다 정확하게 확인할 수 있게 되었다. 이것은 비정상적 부위를 찾아내는 것에 그치지 않고 복잡한 인체 부위에서 정상군과 비정상군을 분리하는데 많은 도움을 주게 되었다. 따라서 임상의는 하나의 검사 영상으로 진단과 치료계획을 동시에 시행할 수 있을 것이다. 또한 감쇠가 잘 되는 흉곽 부위 안에 있는 심근 검사에는, CT로 보다 정확한 감쇠 보정을 할 수 있기 때문에 SPECT 검사 시 관심부위의 관류 상태를 더욱 신뢰할 수 있어 치료 예후의 정당성을 입증할 수 있을 것으로 판단된다.

Keywords