Journal of Industrial Technology (산업기술연구)
- Volume 28 Issue B
- /
- Pages.129-133
- /
- 2008
- /
- 1229-9588(pISSN)
- /
- 1598-1371(eISSN)
Reduction of Seam Line Using an FIR Filter in Spatially Compounded Ultrasonic Diagnostic Images
공간합성된 초음파 의료영상에서 FIR 필터를 이용한 심라인 감소방법
Abstract
A method to reduce seam line artifact in spatial compounding of ultrasonic images is presented. Spatial compounding is a speckle reducing imaging technique in which a number of ultrasound images of a given target that have been obtained from multiple view angles are combined into a single compounded image by combining the data received from each data point in the compounded image. Since different view angle results in different view area, and the images of different view arms are combined into an image, the compounded image consists of regions with different signal to noise ratio, and the boundary lines between these regions are visible as seam lines in the compounded images. In this paper, we present an algorithm that reduces the visibility of this seam line in the spatially compounded images. Design procedure for a FIH filter is described and the results of applying the filter to in-vivo ultrasonic images are analyzed.