Study on a moir$\acute{e}$ Artifact in the Use of Carbon Interspaced Antiscatter Grids for Digital Radiography
탄소 중간물질 그리드를 사용한 DR system에서의 moir$\acute{e}$ artifact에 관한 연구
- Lee, Sung-Ju (Dept. of Radiological Science, Yonsei University) ;
- Cho, Hyo-Sung (Dept. of Radiological Science, Yonsei University) ;
- Choi, Sung-Il (Dept. of Radiological Science, Yonsei University) ;
- Cho, Hee-Moon (Dept. of Radiological Science, Yonsei University) ;
- Oh, Ji-Eun (Dept. of Radiological Science, Yonsei University) ;
- Lee, So-Young (Dept. of Radiological Science, Yonsei University) ;
- Park, Yeon-Ok (Dept. of Radiological Science, Yonsei University) ;
- Lee, Min-Sik (Dept. of Radiological Science, Yonsei University)
- 이성주 (연세대학교 방사선학과) ;
- 조효성 (연세대학교 방사선학과) ;
- 최성일 (연세대학교 방사선학과) ;
- 조희문 (연세대학교 방사선학과) ;
- 오지은 (연세대학교 방사선학과) ;
- 이소영 (연세대학교 방사선학과) ;
- 박연옥 (연세대학교 방사선학과) ;
- 이민식 (연세대학교 방사선학과)
- Published : 2008.12.31
Abstract
Antiscatter grids are widely used in radiography to remove scattered X-rays and thus improve the image contrast. However, the use of grids makes moir
비산란 그리드는 X선 영상에서 산란방사선을 제거시키기 위해 사용되고, 그에 따라 X선 영상의 대조도를 향상 시킬 수 있다. 그러나 많은 경우 디지털 X선 영상에서는 그리드의 부적절한 샘플링으로 인해 moir