Feasibility of Two Dimensional Ion Chamber Array for a Linac Periodic Quality Assurance
선형가속기의 품질관리를 위한 2차원이온전리함배열의 유용성
- Lee, Jeong-Woo (Department of Radiation Oncology, Konkuk University Medical Center, School of Medicine, Konkuk University) ;
- Hong, Se-Mie (Department of Radiation Oncology, Konkuk University Medical Center, School of Medicine, Konkuk University) ;
- Park, Byung-Moon (Department of Radiation Oncology, Konkuk University Medical Center, School of Medicine, Konkuk University) ;
- Kang, Min-Young (Department of Radiation Oncology, Konkuk University Medical Center, School of Medicine, Konkuk University) ;
- Kim, You-Hyun (Department of Radiologic Science, College of Health Science, Korea University) ;
- Suh, Tae-Suk (Department of Biomedical Engineering, School of Medicine, The Catholic University of Korea)
- 이정우 (건국대학교병원 건국대학교 의학전문대학원 방사선종양학교실) ;
- 홍세미 (건국대학교병원 건국대학교 의학전문대학원 방사선종양학교실) ;
- 박병문 (건국대학교병원 건국대학교 의학전문대학원 방사선종양학교실) ;
- 강민영 (건국대학교병원 건국대학교 의학전문대학원 방사선종양학교실) ;
- 김유현 (고려대학교 보건과학대학 방사선학과) ;
- 서태석 (가톨릭대학교 의과대학 의공학교실)
- Published : 2008.06.30
Abstract
Aim of this study is to investigate the feasibility of 2D ion chamber array as a substitute of the water phantom system in a periodic Linac QA. For the feasibility study, a commercial ion chamber matrix was used as a substitute of the water phantom in the measurement for a routine QA beam properties. The device used in this study was the I'm RT MatriXX (Wellhofer Dosimetrie, Germany). The MatriXX consists of a 1,020 vented ion chamber array, arranged in
본 연구의 목적은 방사선치료용 선형가속기의 품질관리를 수행하기 위한 2차원이온전리함배열의 유용성을 검증하기 위함이다. 물팬톰과 필름으로 수행하던 기존의 품질관리 방법을 2차원이온전리함배열(MatriXX, Wellhofer Dosimetrie, Germany)을 이용하여 유용성을 검증하였다. MatriXX는 1,020개의 평판형 전리함(용적: