A Study of Equipment Accuracy and Test Precision in Dual Energy X-ray Absorptiometry
골밀도검사의 올바른 질 관리에 따른 임상적용과 해석 -이중 에너지 방사선 흡수법을 중심으로-
-
Dong, Kyung-Rae
(Department of Radiological Technology, Gwangju Health College) ;
-
Kim, Ho-Sung
(Department of Nuclear Medicine, Asan Medical Center) ;
- Jung, Woon-Kwan (Department of Nuclear Energy Technology, Chosun University)
- Published : 2008.03.31
Abstract
Purpose : Because there is a difference depending on the environment as for an inspection equipment the important part of bone density scan and the precision/accuracy of a tester, the management of quality must be made systematically. The equipment failure caused by overload effect due to the aged equipment and the increase of a patient was made frequently. Thus, the replacement of equipment and additional purchases of new bonedensity equipment caused a compatibility problem in tracking patients. This study wants to know whether the clinical changes of patient's bonedensity can be accurately and precisely reflected when used it compatiblly like the existing equipment after equipment replacement and expansion. Materials and methods : Two equipments of GE Lunar Prodigy Advance(P1 and P2) and the Phantom HOLOGIC Spine Road(HSP) were used to measure equipment precision. Each device scans 20 times so that precision data was acquired from the phantom(Group 1). The precision of a tester was measured by shooting twice the same patient, every 15 members from each of the target equipment in 120 women(average age 48.78, 20-60 years old)(Group 2). In addition, the measurement of the precision of a tester and the cross-calibration data were made by scanning 20 times in each of the equipment using HSP, based on the data obtained from the management of quality using phantom(ASP) every morning (Group 3). The same patient was shot only once in one equipment alternately to make the measurement of the precision of a tester and the cross-calibration data in 120 women(average age 48.78, 20-60 years old)(Group 4). Results : It is steady equipment according to daily Q.C Data with
목적 : 골밀도검사의 중요한 부분을 차지하고 있는 검사장비 및 검사자의 정밀도와 정확도는 환경에 따라 차이가 있기 때문에 질 관리가 체계적으로 이루어져야 한다. 골밀도 검사장비의 노화 및 잦은 고장에 의하여 장비의 교체 및 추가 구입으로 인하여, 추적검사를 하는 환자들의 호환성에 문제가 있다. 따라서 장비 교체 및 증설 후 동일한 장비처럼 호환하여 시용해도 환자의 임상적인 골밀도 변화를 정확하고 정밀하게 반영할수 있는지 알아보고자 한다. 재료 및 방법 : 장비 정밀도는 GE Lunar Prodigy Advance 2 대의 장비 (P1, P2)와 HOLOGIC Spine Phantom(HSP)을 이용하여 각 장비에서 20 번씩 스캔하여 팬텀을 이용한 정밀도 데이터를 획득하였고 (Group 1), 여성 120명 (평균나이 48.78,