Evaluation of Difference between Skin Motion and Tumor Motion for Respiration Gated Radiotherapy
호흡조절방사선치료를 위한 피부움직임과 종양움직임 차이 평가
- Kwon, Kyung-Tae (Department of Radiation Oncology, Asan Medical Center) ;
- Lim, Sang-Wook (Department of Radiation Oncology, Kosin University Gospel Hospital) ;
- Park, Sung-Ho (Department of Radiation Oncology, Asan Medical Center) ;
- Kwon, Soo-Il (Department of Medical Physics, Kyonggi University) ;
- Shin, Sung-Soo (Department of Radiation Oncology, Asan Medical Center) ;
- Lee, Sang-Wook (Department of Radiation Oncology, Asan Medical Center) ;
- Ahn, Seung-Do (Department of Radiation Oncology, Asan Medical Center) ;
- Kim, Jong-Hoon (Department of Radiation Oncology, Asan Medical Center) ;
- Choi, Eun-Kyung (Department of Radiation Oncology, Asan Medical Center)
- 권경태 (서울아산병원 방사선종양학과) ;
- 임상욱 (고신대학교 복음병원 방사선종양학과) ;
- 박성호 (서울아산병원 방사선종양학과) ;
- 권수일 (경기대학교 의학물리학과) ;
- 신성수 (서울아산병원 방사선종양학과) ;
- 이상욱 (서울아산병원 방사선종양학과) ;
- 안승도 (서울아산병원 방사선종양학과) ;
- 김종훈 (서울아산병원 방사선종양학과) ;
- 최은경 (서울아산병원 방사선종양학과)
- Published : 2008.03.31
Abstract
Accounting for tumor motion in treatment planning and delivery is one of the most recent and significant challenges facing radiotherapy. The purpose of this study was to investigate the correlation and clarified the relationship between the motion of an external marker using the Real-Time Position Management (RPM) System and an internal organ motion signal obtained fluoroscope. We enrolled 10 patients with locally advanced lung cancer and liver cancer, retrospectively. The external marker was a plastic box, which is part of the RPM used to track the patient's respiration. We investigated the quantitatively correlation between the motions of an external marker with RPM and internal motion with fluoroscope. The internal fiducial motion is predominant in the caraniocaudal direction, with a range of
본 연구는 호흡조절방사선치료(Respiration gated radiotherapy)를 위해 종양의 실제 움직임과 호흡조절감시장치로 측정한 피부움직임과의 차이를 분석하여 호흡조절방사선치료 시 발생 가능한 오차를 예측하고자 하였다. 호흡에 따른 종양의 움직임을 알아보기 위해 본원에서 2007년1월 10일부터 2월 28일까지 횡경막 주위의 종양 움직임이 큰 폐부위 환자8명, 복부부위 환자 2명에 대해 투시검사기와 호흡조절감시장치를 이용하여 종양의 움직임 영상 및 복부에 위치시킨 적외선 반사장치의 움직임을 측정하였으며, 이 두 측정값의 차이를 정량적으로 비교 평가하였다. 투시검사기에서 종양의 움직임은