Abstract

This research is to develop PPG filtering method for respiration measurement in U-Health Care system. Respiration rate was determined by filtering PPG and analyzing its spectrum. Optimal filter of PPG has been selected to get respiration by testing 120 sets of experiment data using 700 filtering cases. As a result, 2nd order Bessel-filter that used band-pass cutoff frequency at 0.175~0.4Hz with second order was good at developing respiration signal. Respiration signal in time domain could be continuously analyzed by converting frequency domain using spectrum analysis. 24 seconds has been found to be optimal time duration of collecting PPG data for determining respiration. Therefore, this study was successful of getting not only heart activity but also respiration by only PPG. Minimal invasive measurement obtaining multi-bio information by one sensor can be expected to apply to U-Health Care and human computer interaction.

Keywords

References

  1. 이창관, 입원 호흡재활 프로그램이 만성 폐질환자의 호흡곤란, 운동능력과 건강관련 삶의 질에 미치는 효과, 대한간호학회지, 37(3), 343-352, 2007 https://doi.org/10.4040/jkan.2007.37.3.343
  2. 우종민, 심박동수 변이 측정법의 개념과 임상적 활용, 정신신체의학, 12(1), 3-14, 2004
  3. Allen, J. and Murray, A., Variability of photoplethysmography peripheral pulse measurements at the ears, thumbs, and toes, IEEE Proc. Sci. Meas. Technol, 147, 403-407, 2000 https://doi.org/10.1049/ip-smt:20000846
  4. Barschdorff., and Zhang., Respiratory Rhythm Detection with Photoplethysmographic Methods, Medicine and Biology Society: Proc. 16th Ann. Int. Conf. of the IEEE, 912-913, 1994
  5. Emily, A. B., Frank H. W. and James J. G., Respiratory sinus arrhythmia, emotion and emotion regulation during social interaction, Psychophysiology, 43(6), 612-622, 2006 https://doi.org/10.1111/j.1469-8986.2006.00467.x
  6. Johansson, A., Öberg, P. Å. and Sedin, G., Monitoring of heart and respiratory rates in newborn infants using a new photoplethysmographic technique, Journal of Clinical Monitoring and Computing, 15, 461-467, 1999 https://doi.org/10.1023/A:1009912831366
  7. Johansson, A. and Öberg, P. Å., Estimation of respiratory volumes from the photoplethysmographic signal. Part 1: experimental results, Journal of Clinical Monitoring and Computing, 37, 42-47, 1999a
  8. Johansson, A., Neural network for photoplethysmographic respiratory rate monitoring, Journal of Clinical Monitoring and Computing, 41, 242-248, 2003
  9. Kim, M. K., Park, J. H. and Jee, K. Y., Demand analysis and market strategies of u-health, advanced communication technology The 9th international conference on, 1, 102-105, 2007
  10. Lindberg, L. G., Ugnell, H. and Öberg, P. Å., Monitoring of Respiratory and Heart-Rates Using a Fiberoptic Sensor, Med. Biol. Eng. Comput, 30, 533-537, 1992 https://doi.org/10.1007/BF02457833
  11. Nilsson, L., Johansson, A. and Kalman, S., Monitoring of respiratory rate in postoperative care using a new photoplethysmographic technique, Journal of Clinical Monitoring and Computing, 16, 309-315, 2000 https://doi.org/10.1023/A:1011424732717
  12. Nilsson, L., Johansson, A. and Kalman, S., Respiratory variations in the reflection mode photoplethysmographic signal. Relationships to peripheral venous pressure, Medical & Biological Engineering & Computing, 41, 249-254, 2003a https://doi.org/10.1007/BF02348428
  13. Nilsson, L., Johansson, A. and Kalman, S., Macrocirculation is not the sole determinant of respiratory induced variations in the reflection mode photoplethysmographic signal, Physiol Meas., 24, 925-937, 2003b https://doi.org/10.1088/0967-3334/24/4/009
  14. Nilsson, L., Johansson, A. and Kalman, S., Respiration can be monitored by photoplethysmography with high sensitivity and specificity regardless of anaesthesia and ventilatory mode Acta Anaesthesiol, Anaesthesiologica Scandinavica, 49(8), 1157-1162, 2005 https://doi.org/10.1111/j.1399-6576.2005.00721.x
  15. Yoon, G., Lee, J. Y., Jeon, K. J., Park, K. K. and Kim, H. S., Development of a compact home health monitor for telemedicine, Telemed. J. E. Health, 11, 660-667, 2005 https://doi.org/10.1089/tmj.2005.11.660

Cited by

  1. Design of ICA to Extract Respiration Signal From PPG Signal vol.9, pp.2, 2011, https://doi.org/10.6109/jicce.2011.9.2.220